8.4  Accessories

The On-X Prosthetic Heart Valve is designed to be used 
only with On-XLTI On-X instruments. The instruments, 
supplied separately, are provided as a set, which 
includes sizers, rotators, an instrument handle, and a 
leaflet probe. The instruments are reusable.

CAUTION: Sizers and instrument handles have 
metallic regions that are bendable. Repeated 
bending of these metallic regions can lead to fatigue 
and fracture. To avoid instrument fracture during 
use, the stem should be inspected for surface 
cracks before and after each time it is bent. If metal 
fatigue surface cracks are present, the sizer and/
or instrument handle should be discarded and 
replaced. Contact On-XLTI Customer Service to order 
replacements.
CAUTION: Leaflet probes and rotators are flexible, 
but are not intended to be bent to a permanently 
deformed state.

Sizer

The sizer is used to gauge the resulting tissue annulus 
diameter after the annulus is prepared for implant. The 
sizer has a bendable stem on each end. The sizers are 
cylindrical for size 19 mm through 25 mm valves and 
conical for size 27/29 mm and 31/33 mm valves (Figure 3a 
and 3b). To facilitate sizer selection, refer to Table 2.

Sizer Choice

Position of sewing ring

Size

Valve Type

Sizer Type

Use Replicate Sizer

19

Aortic

Cylindrical

YES

Intrasupra-annular

21

Aortic

Cylindrical

YES

Intrasupra-annular

23

Aortic

Cylindrical

YES

Intrasupra-annular

25

Aortic

Cylindrical

YES

Intrasupra-annular

27/29

Aortic

Conical

YES

Intra-annular

19*

Aortic Conform-X

Cylindrical

YES

Intrasupra-annular

21*

Aortic Conform-X

Cylindrical

YES

Intrasupra-annular

23*

Aortic Conform-X

Cylindrical

YES

Intrasupra-annular

25*

Aortic Conform-X

Cylindrical

YES

Intrasupra-annular

27/29*

Aortic Conform-X

Conical

YES

Intra-annular

19*

Aortic Anatomic

Cylindrical

YES

Intrasupra-annular

21*

Aortic Anatomic

Cylindrical

YES

Intrasupra-annular

23*

Aortic Anatomic

Cylindrical

YES

Intrasupra-annular

25*

Aortic Anatomic

Cylindrical

YES

Intrasupra-annular

27/29*

Aortic Anatomic

Conical

YES

Intra-annular

23*

Mitral

Cylindrical

NO

Supra-annular

25

Mitral

Cylindrical

NO

Supra-annular

27/29

Mitral

Conical

NO

Supra-annular

31/33

Mitral

Conical

NO

Supra-annular

25/33

Mitral Conform-X

Cylindrical or Conical

NO

Supra-annular

Replica Sizers

Aortic replica sizers are provided for all aortic valve 
sizes (Figure 3a). They model the On-X standard aortic 
valve profile. They are used after sizing for standard, 
Conform-X, and Anatomic sewing ring configurations 
to assure fit of the aortic valve without obstruction of 
the coronary arteries. The size 19 through 25 aortic 
replica sizers shape is intended to model intrasupra-
annular positioning. The size 27/29 aortic replica sizer is 
intended to model intra-annular positioning.

Table 2: Sizer Selections - Regardless of Aortic Holder Type

Figure 3a: Sizer and Replica Sizer

Figure 3b: Sizer 

Cylindrical

Aortic Replica

Conical

* Not available in all markets

Summary of Contents for On-X Aortic Heart Valve and Extended Holder

Page 1: ...eart Valve with Conform X Sewing Ring REF ONXAC On X Mitral Heart Valve with Conform X Sewing Ring REF ONXMC On X Aortic Heart Valve with Anatomic Sewing Ring REF ONXAN On X Aortic Heart Valve and Ext...

Page 2: ...SSMENT AND VALVE ROTATION 11 9 7 VALVE ORIENTATION 12 10 POSTOPERATIVE INFORMATION 12 10 1 MAGNETIC RESONANCE IMAGING MRI COMPATIBILITY 12 10 2 RETURNED GOODS 12 11 PATIENT INFORMATION 12 11 1 PATIENT...

Page 3: ...eplica sizers verify the aortic valve 10 Figure 12 Supra annular valve positioning 11 Figure 13 Sewing ring cross sections 11 Figure 14 Removing the valve holder 11 Figure 15 Insert valve rotator 12 F...

Page 4: ...a travel arc of 50 to the closed position The orifice is composed of graphite substrate coated with On X Carbon a pure unalloyed form of pyrolytic carbon The leaflets consist of On X Carbon deposited...

Page 5: ...esis regurgitation prosthesis structural dysfunction prosthesis thrombosis stroke thromboembolism It is possible that these complications could lead to reoperation explantation permanent disability de...

Page 6: ...rded on the outer package label Appropriate inventory control should be maintained so that prostheses with earlier expiration dates are preferentially implanted and expiration is avoided To protect th...

Page 7: ...YES Intrasupra annular 23 Aortic Cylindrical YES Intrasupra annular 25 Aortic Cylindrical YES Intrasupra annular 27 29 Aortic Conical YES Intra annular 19 Aortic Conform X Cylindrical YES Intrasupra...

Page 8: ...n and Resterilization The On X Prosthetic Heart Valve is provided sterile If the sterility expiration date has passed or if upon removal from the outer box the valve container is damaged or the steril...

Page 9: ...the patient or place it in the patient s records Figure 7a Twist off outer lid Figure 7b Remove by pull tab Figure 7c or invert on sterile field 4 Open the outer container Twist off outer lid package...

Page 10: ...thetic Heart Valve sizers when sizing the annulus Sizers contain cylindrical conical and aortic replica ends Refer to Table 2 to facilitate sizer selection Cylindrical sizers correspond to the valve s...

Page 11: ...mplanted using a pledgetted or non pledgetted everting mattress suture technique although non everting and continuous suture techniques have also been used with success CAUTION When seating the valve...

Page 12: ...e to rotation is encountered The torque required to rotate the valve in situ should be about the same as that required when testing rotation before implantation If noticeably greater torque is require...

Page 13: ...mplete the card and carry it with them at all times 11 3 Patient Information Booklet On XLTI has made available a patient information booklet that the physician may choose to provide to the patient pr...

Page 14: ...s in an international multicenter prospective non randomized study with retrospective controls Patients requiring isolated mitral heart valve replacement were enrolled from 1996 to 2001 at 16 centers...

Page 15: ...d as linearized rates based on total patient years 3 Freedom from event was calculated based on the method of Kaplan Meier SE Standard Error 4 n number of patients in each category N total number of s...

Page 16: ...3 0 7 99 0 0 7 99 0 0 7 Explant 1 0 4 3 0 7 98 0 1 0 98 0 1 0 Hemolysis5 0 0 0 0 0 0 100 0 0 100 0 0 Hemorrhage6 all 4 1 8 6 1 4 96 4 1 3 94 4 2 0 Hemorrhage major 4 1 8 2 0 5 97 0 1 2 97 0 1 2 Periv...

Page 17: ...tis 18 7 8 Gastroinestinal Ulcer 18 7 8 Chronic Kidney Failure 13 5 6 Carotid Artery Disease 12 5 2 Coronary Artery Bypass Graft 10 4 4 Cancer 10 4 4 Previous Mitral Valve Replacement 9 3 9 Cardiomyop...

Page 18: ...ts followed reproduced from Figure 2 Nd number of patients for which NYHA data were collected not including missing 3 n number of patients in each category 4 Undetermined means data were collected but...

Page 19: ...2 0 4 3 Regurgitation Nd 20 Nd 37 Nd 68 Nd 36 Nd 25 n n Nd n n Nd n n Nd N n Nd n n Nd 0 5 25 0 9 24 3 27 39 7 17 47 2 7 28 0 1 2 12 60 0 25 67 6 37 54 4 16 44 4 17 68 0 3 2 10 0 0 0 0 3 4 4 2 5 6 1 4...

Page 20: ...Bldg B Austin Texas 78752 U S A TEL 512 339 8000 FAX 512 339 3636 WEB www onxlti com EMAIL onx onxlti com CryoLife Europa Ltd Bramley House The Guildway Old Portsmouth Road Guildford Surrey GU3 1LR Un...

Reviews: