FUSE REPLACEMENT
WARNING: Before replacing a fuse, you must first determine the reason why it has blown and carry out
the replacement only once the causes have been eliminated. The new safety fuse must have the same
features as the replaced fuse. These specifications are shown in the electrical diagram of the equipment.
On the basis of the dental chair is easy to spot the General protection fuse (B) external-replaceable with
a screwdriver; before the replacement make sure that the main switch (A) is switched off. Fuse
replacement should occur only after you remove the cause that caused the intervention; If the problem
persists after replacing the fuse you must request the assistance of a OMS qualified technician.
The fuses are also present on the electronic cards;
The replacement of fuses placed on electronic boards (placed inside the casing at the base of the dental
chair) shall be carried out only by an authorised O.M.S..
Are stated below, for illustrative purposes only, the values of the main fuses:
B.
General fuse 8AT, 230 Vac