DUKE EASY S.T.01/3AW Rev.0 01/2017
17
The spray is always preselected. To adjust the water to spray one acts on the faucet in place behind the
Chair. By turning the tap clockwise you get a gradual decrease of water flow to the spray until its complete
exclusion, the opposite rotating in the opposite direction. It is recommended not to intervene in other
regulations placed at the side of the spray faucet.
If the instrument is equipped with optical fibres lighting system, these are switched on by pressing the
control
command pad spot Tablet (when the function is on, it affects all instruments). The
activation of the function is displayed by the lit led (D).
Air pressure Recommended (measured during operation of the turbine) is set at the time of testing in
the O.M.S.. Calibration must be re-evaluated at the time of installation of the appliance using a
manometer, respecting the pressure values indicated by the manufacturer of the turbine. This must be
done by an authorised O.M.S..
For more details see the attached manufacturer's instructions.
ELECTRIC MICROMOTOR MODULE
The function of the micromotor is achieved by lifting the instrument from its place, moving it to the
surgical field and then pressing the foot pedal (see operation section MICROMOTOR).
Lifting the instrument from the table are inhibited all other dynamic tools and all the movements of the
dental chair.
The micromotor allows you to work with a variable speed from a minimum of about 900 RPM to a maximum
of 40,000 rpm
Normally the micromotor will have a clockwise direction rotation. For reversal of direction of rotation,
after lifting the tool, activate the reverse command
in the instruments table.
It is recommended that the change of direction of rotation is made only when the motor is stopped, to
prevent damage.
The spray is preset by typing the command
; the activation of the function by the lit led is
displayed. See also section OPERATION micromotor.
To adjust the water to spray one acts on the faucet in place behind the Chair. By turning the tap clockwise
you get a gradual decrease of water flow to the spray until its complete exclusion, the opposite rotating
in the opposite direction. It is recommended not to intervene in other regulations placed at the side of
the spray faucet.