K3TS PANEL METER SETTINGS
Level 3
One linear Sensor only. IN A is used.
No previous average comparison.
Peak-to-peak hold.
Level 2
±
9.999 V is selected as the input mode.
The number of measuring operations used to obtain the average.
Timing delay setting. Set to 0.
Level 1
to
Select HH, H, L, or LL for determining the eccentricity of an object when the comparison output is used.
Converts the -4 to 4V output of the Z4M into the actual dimension.
If the Z4M-W40 is used, the linear output will increase by 0.4 V with a displacement value of
1 mm. The output will be 0 V when the object is in the center of the measurement range
(40 mm between the object and the Sensor) and the output will be 0.4 V when the distance between the
object and the Sensor is 41 mm. The output value is used for scaling.
X2:
When the linear output is 0V.
Y2:
A distance of 40 mm between the object and the Sensor is regarded as the standard distance (0 mm).
X1:
The linear output is 0.4 V (400 mV).
Y1:
When the sensing distance is 41 mm, the difference between the sensing distance and the standard
distance is 1 mm.
Set the decimal points of Y2 and Y1.
Prohibits comparison set point change in the RUN mode. After all settings have been completed, set to
ADJUSTING/OPERATING
Axis and Distance Adjustment
1.
The Z4M uses an infrared laser beam. The spot of the beam, however,
can be confirmed in the dark by locating blue paper at the measurement
point so that the beam will be reflected by the blue paper. The Sensor
incorporates a laser beam delaying circuit, so the laser will start emitting
3 to 10 s after the Sensor is turned on.
2.
Ensure that the center of the object displacement will be in the center of
the Sensor’s measurement range.
Both the FAR and NEAR
indicators are lit when the
sensing object is at the
measurement point.
Z4M