background image

Flat-blade
screwdriver

Offset adjustor

Span adjustor

Connect to an Intelligent
Panel Meter such as a
K3TX or K3TS.

Offset Adjustment

Offset adjustment is made using a white ceramic object before shipping,
so the user may not need to adjust the output.

To set the Sensor output to 0 V at the standard distance, connect an
Intelligent Panel Meter (such as K3TX or K3TS) to the output terminal of
the Sensor. Place a standard object in front of the Sensor and adjust the
output with the offset adjuster.

Span Adjustment

Span adjustment is made using a white ceramic object before shipping, so
the user may not need to adjust the span.

To calibrate the displacement output, adjust the span as follows.

1.

Locate the object at the standard distance and adjust the offset ad-
justment.

2.

Move the object for a specified distance and adjust the span adjuster
so that the voltage will change properly according to the displace-
ment.

Follow these steps to convert

±

4 V to

±

5V.

1.

Connect an Intelligent Panel Meter to the Sensor and move the object
so that the output will be 4 V.

2.

Use the span adjuster to adjust the voltage to 5 V.

Dimensions

Unit: mm (inch)

SENSOR: Z4M-S40 (FOR Z4M-W40), Z4M-S100 (FOR Z4M-W100)

Mounting Holes

Lens
(7 dia.)

Front face

Two 4.5-dia. mounting
holes

Measurement point

Vinyl-insulated round cable,
6.6 dia, standard length: 2 m

Connector

Emitting
axis

17

10.5

1.5

32

43

39

6.75

3.5

6.75

44

11

3.5

6.75

37

30

14.7 dia.

*

23

°

for Z4M-S40 and 11

°

for Z4M-S100.

**

40mm (1.6in) for Z4M-S40 and 100mm (3.9in) for Z4M-S100.

6.75

**

Two M4 Taps

37

±

0.2

39

±

0.2

44

(1.73)

(1.54)

(1.46)

17.6
(0.69)

46
(1.81)
26

*

Z4M

Z4M

Summary of Contents for Z4M-W10

Page 1: ... minimizes sensing errors caused by color or surface variations Compact sensor head incorporates easy to use range indicators Amplifier can be DIN rail mounted Connects to the Z49 SF2 Laser Safety Kit to conform to FDA standards Ordering Information SENSORS Sensing distance Resolution response speed selected Part number 40 10 mm 1 57 40in 1 5 µm 60 ms 10 µm 2 ms 40 µm 0 15 ms Z4M W40 100 40 mm 3 9...

Page 2: ...onversion value of the analog output Condition white alumina ceramic at the measurement point Linearity see Note 2 1 FS see Note 1 1 5 FS see Note 1 Response time see Note 3 0 15 ms 2 ms 60 ms switch selectable 0 7 ms 20 ms 500 ms switch selectable Sensitivity selector WHITE BLACK AUTO switch selectable Temperature characteristics at measurement point Sensor 0 03 FS C Amplifier 0 03 FS C Sensor 0 ...

Page 3: ...sponse time Response time CHARACTERISTICS Power supply voltage 12 to 24 VDC 10 ripple p p 10 max Current consumption 120 mA max Dielectric strength Sensor 1 000 VAC 50 60 Hz for 1 min Amplifier 300 VAC 50 60 Hz for 1 min Vibration resistance Destruction 10 to 55 Hz 1 5 mm double amplitude for 32 minutes each in X Y and Z directions Shock resistance Destruction 300 m s2 30G for 3 times each in X Y ...

Page 4: ...or FS White ceramic Black paper Rolled steel plate Inclination angle Inclination angle Inclination angle Error FS See note Note Rolled steel cannot be meas ured due to excessive light reflection SPOT DIAMETER TYPICAL EXAMPLE L l1 l2 Cross section of the beam Z4M W40 Distance L 30 mm 1 18 in 40 mm 1 57 in 50 mm 1 97 in 1 1 2 mm 0 047 in 0 6 mm 0 024 in 0 2 mm 0 008 in 2 0 6 mm 0 024 in 0 3 mm 0 012...

Page 5: ...surement point distance in mm Error FS White ceramic Black paper Rolled steel plate Inclination angle 10 Z4M W100 Inclined Object Inclined Object Inclined Object Inclination angle 0 Error FS Measurement point distance in mm White ceramic Black paper Rolled steel plate White ceramic Black paper Rolled steel plate Inclination angle 10 Error FS Measurement point distance in mm White ceramic Black pap...

Page 6: ...An English warning sticker is provided with the model At tach the sticker here Emitter laser beam emitter Receiver Connector AMPLIFIER Stability indicator Offset adjustor Span adjustor Sensitivity selector Response speed selector Connector terminals NEAR indicator FAR indicator Z4M Z4M ...

Page 7: ...is stable If the indicator is not lit in green while the sensitivity selector is set to WHITE set it to BLACK or AUTO for a more stable measuring operation Normal output can be obtained even if the indicator is not lit in green The indicator will be lit in red when there is no object in front of the Sensor or if the light reflected from the object is either insufficient or too intense If the indic...

Page 8: ...V 60 to 140 mm Z4M W100 Output impedance 100 Ω typical The linear output will be locked between 6 and 8 V when the enable output is OFF Laser OFF input Amplifier The laser OFF input controls laser emission Laser emission is turned on when there is no laser OFF input with a current leakage of 0 1 mA max and it is turned off when the laser OFF input is short circuited to the 0 V terminal with a resi...

Page 9: ...e linear output is 0V Y2 A distance of 40 mm between the object and the Sensor is regarded as the standard distance 0 mm X1 The linear output is 0 4 V 400 mV Y1 When the sensing distance is 41 mm the difference between the sensing distance and the standard distance is 1 mm Set the decimal points of Y2 and Y1 Prohibits comparison set point change in the RUN mode After all settings have been complet...

Page 10: ...ws 1 Locate the object at the standard distance and adjust the offset ad justment 2 Move the object for a specified distance and adjust the span adjuster so that the voltage will change properly according to the displace ment Follow these steps to convert 4 V to 5V 1 Connect an Intelligent Panel Meter to the Sensor and move the object so that the output will be 4 V 2 Use the span adjuster to adjus...

Page 11: ...length 2 m Enlarged view of mounting hole 25 Two M4 Taps 25 30 1 18 80 98 3 86 89 15 26 4 8 42 7 5 2 12 5 3 3 6 21 35 2 55 2 17 6 0 7 4 4 5 5 30 Mounting Holes 1 39 3 15 3 50 ACCESSORIES SOLD SEPARATELY Oil resistive vinyl insulated shielded cable external diameter 6 6 11 9 dia 14 dia 14 7 dia 9 5 dia 43 43 3 m 9 84ft or 8 m 26 25 ft Extension Cable Z49 C1 Laser Safety Kit Z49 SF2 Z4M Z4M ...

Page 12: ...ut of the Sensor will be on hold Linear output 4 to 4 V will be output according to the displacement value Wire the Sensor correctly or the sensor may be damaged The linear output line must not be in contact with any other line Linear GND Used as a ground 0V terminal for the linear output of the Sensor Connect this line to the input device Note 1 The 0 V blue and linear GND shielded line are inter...

Page 13: ...er than 130 Vrms between terminals Laser Radiation Emission Indicator The system incorporates an emission indicator which gives a visible or audible signal from 2 to 20 seconds before and during radiation emission Visible indicators should be visible to users wearing laser protection glasses Safety Interlocks Safety interlocks are fitted to each part of the protective housing to prevent the emissi...

Page 14: ...placement Sensor cannot accurately detect objects that are mirror like transparent inclined or have an extremely small reflection ratio smaller than the Sensor s sensing spot diameter DANGER Do not wire the power supply cable for the Z4M Sensor in the same conduit with high voltage lines or power lines The result would be interference damage or malfunction WIRING A Z49 C1 Extension Cable can be co...