31
Product Description
Electronic Nerve Stimulator
Model (code)
HV-F013(HV-F013-AP)
Electrical Characteristics
3V (with two alkaline
“AAA”
batteries [LR03] or manganese [R03])
Battery Life
New batteries (two AAA alkaline
batteries) will last for approx. 3 months
(when used for 15 minutes a day, Lower
Back Mode, max. intensity).
Energy Consumption
Approx. 30mA
Frequency
Approx. 1 to 108Hz
PULSE Duration
100 μsec
Maximum Output Voltage
32V (during 500Ω load)
Power Control
10 intensity levels
Operating Temperature,
Humidity (When using product)
10 °C to 40 °C, 30 to 80% RH
Storage Temperature, Humidity 0 °C to 40 °C, 30 to 80% RH
Transportation Temperature,
Humidity, Air Pressure
-20 °C to 60 °C, 10 to 95% RH,
700 to 1060 hPa
Weight
Approx. 75g (incl. batteries)
Outer Dimension
(Width × Height × Depth)
Approx. 75 × 70 × 22mm
Classification of ME equipment
Internally powered
IP classification
IP 22*
Contents
Main Unit, Electrode Cord, Long
Life Pads, Pad Holder, 2 “AAA”
Batteries(LR03), Instruction Manual,
Quick Start Guide/Pad Placement Guide
This OMRON product is produced under the strict quality system
of OMRON HEALTHCARE Co. Ltd., Japan.
Designed for a minimum of 5 years life expectancy.
*
IP classification is degrees of protection provided by enclosures
in accordance with IEC 60529. This device is protected against
solid foreign objects of 12.5 mm diameter and greater such as
a finger. This device is protected against oblique falling water
drops which may cause issues during a normal operation.
NOTE:
These specifications are subject to change.