2.1 Inserting/replacing batteries
When should users change the batteries?
1. When the symbol shows on the display, it means the battery
power is low and need to replace batteries as soon as possible.
Even the battery power is low, HGM-111 meter still can work well:
You still can use HGM-111 and the test results will still be accurate,
but we strongly suggest you replace the batteries as soon as
possible.
2. There is not enough power to support HGM-111 to operate when the
symbol is shown on display after powering on. Thus, the
meter will shut down after 20 seconds when the symbol is shown
on display after powering on. It represents that the battery power is
too low to perform a test. You have to replace the batteries to
continue using the HGM-111 meter. Time and date must be reset
after replacing the batteries, but the results records still remain.
How to replace HGM-111’s batteries
1. Turn HGM-111 to the back side and open the battery compartment.
2. Take out the batteries and insert two CR2032 batteries with the (+)
sign facing up.
3. Push back the battery compartment until hearing a “click” sound.
4. Try to turn on the device to see whether the HGM-111 can be
powered on. If not, please check the above steps.
5. After turning on the HGM-111 meter, it will enter into Buzzer and date
setting mode directly. (Please refer to setting up your HGM-111 of
2.2 for relative settings)
Chapter 2 Installing and Setting up the HGM-111 meter
6
BLOOD GLUCOSE MONITORING SYSTEM-HGM-111