Cleaning and disinfection

Disinfect the nebuliser before first use and at least once per day when in use.
Procedure:
Disassemble the nebuliser kit by disconnecting the air tube and mouthpiece or mask.
Remove the nebuliser cap from the medication tank and remove the vaporizer head.
Also refer to the illustrations as shown under step 3 and 4 (page 5).
Disinfect all disassembled nebuliser parts –NOT the air tube and masks- by placing them in boiling water
(max.10 minutes) or in an autoclave (max. 121ºC, refer to the autoclave instructions for the duration).

To clean the nebuliser and mask immediately after each treatment, it is sufficient to rinse the disassembled
parts under running warm tap water.
The masks can be disinfected by any commercially available disinfectant, containing hypochloride.
After cleaning/disinfecting of the parts, place them on a clean (paper) towel and let them air dry.

For hygienic reasons, each person should use their own nebuliser kit. It is recommended that the nebuliser
kit is replaced every 6 months (or earlier in case the nebuliser is obstructed) in order to maintain efficient
inhalation. Contact your OMRON dealer for replacements.

The exterior of both the air tube and compressor unit should be cleaned once daily with a clean, damp 
piece of cloth. Do not use any volatile fluids. Never submerge the compressor unit in water!
If there is some moisture in the air tube, connect the air tube to the compressor unit and leave the other 
end open. Then switch on the compressor unit and blow the tube dry.

The exterior of both the air tube and compressor unit should be cleaned once daily with a clean, damp 
piece of cloth. Do not use any volatile fluids. Never submerge the compressor unit in water!
If there is some moisture in the air tube, connect the air tube to the compressor unit and leave the other 
end open. Then switch on the compressor unit and blow the tube dry.

Storage and maintenance

Put the nebuliser kit, masks and airtube in a clean bag and place them in the storage pocket of the car-
rying case.

Wrap the mains cable and place it in its storage compartment at the rear of the compressor unit.

Do not store the compressor on the floor.

The unit does not contain any user serviceable parts and does not require oil. Do not attempt to 
lubricate any internal parts. Do not open the compressor unit.

In case you encounter difficulties with your nebuliser, please contact your local OMRON dealer.
(address on/inside package).

Disposal of this product should be carried out in accordance with the national regulations for the 
disposal of electronic products.

English

7

Summary of Contents for CompAIR CXpro

Page 1: ...ct You can be sure that you have made a wise choice By buying the OMRON CompAIR CXpro Nebuliser you have chosen a high quality innovative instrument During its development particular attention was given to reliability convenient and easy handling Before using the nebuliser for the first time please read through this manual carefully If you still have any questions regarding its use please contact ...

Page 2: ... The unit does not contain any user serviceable parts and does not require oil Do not attempt to lubricate any internal parts Never open the unit Make sure that the electrical rating shown on the rating label on the bottom of the unit corresponds to your mains voltage and frequency before plugging in the nebuliser Always disconnect the nebuliser from the mains supply before you start cleaning it T...

Page 3: ...mpAIR CXpro Nebuliser 1 Nebuliser kit JetAIRplus see also on page 4 2 Air tube 3 Air filter compartment air inlet 4 Air outlet connection 5 Power OFF ON switch O I 6 Compressor unit 7 Air ventilation holes English 3 1 7 6 2 3 4 5 ...

Page 4: ...buliser kit JetAIRplus 8 Valve cap 9 Exhalation valve 10 Mouthpiece 11 Inhalation valve 12 Nebuliser cap 13 Vaporiser head 14 Jet nozzle inside medication tank 15 Medication tank English 4 1 2 3 4 5 6 9 11 12 13 14 8 10 15 ...

Page 5: ...to the OFF O position and plug in the mains cable For steps 3 7 refer to the drawing below 3 Remove the nebuliser cap from the medication tank by turning it counterclockwise and lifting it Remove the mouthpiece 4 Remove the vaporiser head 5 Fill the medication tank with the prescribed amount and type of medication max 6 ml 6 Reinsert the vaporisation head 7 Reconnect the nebuliser cap to the medic...

Page 6: ... are automatically controlled by your breathing During inhalation air will enter the nebuliser through the inha lation valve During your exhalation exhaled air will leave the nebuliser through the exhalation valve Do not try to breathe too quickly Pause when you feel like having a rest Breathe in Breathe out 11 When the treatment is finished set the OFF ON switch O I to the OFF O position and disc...

Page 7: ...and compressor unit should be cleaned once daily with a clean damp piece of cloth Do not use any volatile fluids Never submerge the compressor unit in water If there is some moisture in the air tube connect the air tube to the compressor unit and leave the other end open Then switch on the compressor unit and blow the tube dry The exterior of both the air tube and compressor unit should be cleaned...

Page 8: ...l socket Make sure that the ends of the air tube are fitted tightly onto the compressor unit and onto the medication tank Check whether the medication tank is empty If so and if necessary refill the medication tank Is the vaporiser head installed correctly in the medication tank If not install it and press it down into the jet nozzle Check whether the air tube is twisted If so untwist Check whethe...

Page 9: ...H Storage conditions Temperature 25 C 70 C Humidity 10 95 RH Supply voltage AC 230 V 50 Hz Power consumption 85 W Electrical safety Fuse T2 5AL250V UR Rit Class 2 Type B Contents of the package Compressor unit nebuliser kit JetAIRplus air tube flexible child mask flexible adult mask carrying case 5 spare filters instruction manual guarantee card value measured with 0 9 NaCl solution Please note th...

Reviews: