General Information

The OMRON CompAIR CX

pro

is a nebuliser for the inhalation of medical aerosols to treat asthma, chronic

bronchitis and other respiratory tract disorders. This nebuliser is suitable for all commonly used inhalation
solutions. Use only the type and amount of medication prescribed by your doctor.

The advanced nebulisation system of the OMRON CompAIR CX

pro

produces a very low particle size distri-

bution and therefore a medication vapour that will reach effectively and reliably the upper and lower respirato-
ry passages.Due to the inhalation and exhalation valves, the amount of medication deposited into the lungs
is maximized and medication waste is minimized.
The OMRON CompAIR CX

pro

is suitable for inhalation at home, in the hospital or at a doctor's practice.

Safety precautions

Please read the following safety precautions carefully, as they contain important information concerning safe
set up, use and maintenance of your nebuliser.

Keep this manual for future reference.

As soon as you have unpacked the nebuliser, check it for visible damages or defects.
In particular check for cracks in the plastic housing, which could expose electrical components. If this 
is the case or if you are in doubt, contact your OMRON dealer directly.

Only use the nebuliser for its intended purpose; for inhalation therapy. Any other form of use constitutes
an improper use and is therefore dangerous. The manufacturer cannot be held liable 
for any damage caused by improper or incorrect use.

Do not leave the unit unattended with people who require supervision (e.g. people with infirmities or
children).

The unit does not contain any user serviceable parts and does not require oil. Do not attempt to 
lubricate any internal parts. Never open the unit.

Make sure that the electrical rating shown on the rating label on the bottom of the unit corresponds to
your mains voltage and frequency before plugging in the nebuliser.

Always disconnect the nebuliser from the mains supply before you start cleaning it.

The unit is not protected against water penetration, so: Never touch the unit when your hands are wet 
or moist. Do not use the nebuliser while you are in the bath or having a shower.
Never submerge the nebuliser in water. In case this occurs accidentally, always first disconnect the 
power supply plug from the wall socket, before reaching for the unit. After having removed the unit from
the water, do not use it again. Contact your OMRON dealer and have your unit checked.

Always switch the unit off first, before you disconnect the power plug.

Do not use the unit in the presence of anaeststhetic mixtures which are inflammable when they come
into contact with air, oxygen or nitrous oxide.

Keep the power cable away from heated or hot surfaces.

If you discover that the nebuliser or its mains cable is damaged or does not operate properly, switch it 
off immediately, disconnect the power plug and stop using it. Contact your OMRON dealer and have your
unit checked.

Always use original OMRON nebuliser kit and air filter; order no. for replacements can be found in this
manual.

Do not block the air ventilation holes on left/right side of the compressor unit.

Place the compressor unit on a flat and stable surface during operation.

Although this device fulfils the provisions of the EMC (Electromagnetic Compatibility) directive, the use of
it should be avoided in direct vicinity of other electric devices.

English

2

Summary of Contents for CompAIR CXpro

Page 1: ...ct You can be sure that you have made a wise choice By buying the OMRON CompAIR CXpro Nebuliser you have chosen a high quality innovative instrument During its development particular attention was given to reliability convenient and easy handling Before using the nebuliser for the first time please read through this manual carefully If you still have any questions regarding its use please contact ...

Page 2: ... The unit does not contain any user serviceable parts and does not require oil Do not attempt to lubricate any internal parts Never open the unit Make sure that the electrical rating shown on the rating label on the bottom of the unit corresponds to your mains voltage and frequency before plugging in the nebuliser Always disconnect the nebuliser from the mains supply before you start cleaning it T...

Page 3: ...mpAIR CXpro Nebuliser 1 Nebuliser kit JetAIRplus see also on page 4 2 Air tube 3 Air filter compartment air inlet 4 Air outlet connection 5 Power OFF ON switch O I 6 Compressor unit 7 Air ventilation holes English 3 1 7 6 2 3 4 5 ...

Page 4: ...buliser kit JetAIRplus 8 Valve cap 9 Exhalation valve 10 Mouthpiece 11 Inhalation valve 12 Nebuliser cap 13 Vaporiser head 14 Jet nozzle inside medication tank 15 Medication tank English 4 1 2 3 4 5 6 9 11 12 13 14 8 10 15 ...

Page 5: ...to the OFF O position and plug in the mains cable For steps 3 7 refer to the drawing below 3 Remove the nebuliser cap from the medication tank by turning it counterclockwise and lifting it Remove the mouthpiece 4 Remove the vaporiser head 5 Fill the medication tank with the prescribed amount and type of medication max 6 ml 6 Reinsert the vaporisation head 7 Reconnect the nebuliser cap to the medic...

Page 6: ... are automatically controlled by your breathing During inhalation air will enter the nebuliser through the inha lation valve During your exhalation exhaled air will leave the nebuliser through the exhalation valve Do not try to breathe too quickly Pause when you feel like having a rest Breathe in Breathe out 11 When the treatment is finished set the OFF ON switch O I to the OFF O position and disc...

Page 7: ...and compressor unit should be cleaned once daily with a clean damp piece of cloth Do not use any volatile fluids Never submerge the compressor unit in water If there is some moisture in the air tube connect the air tube to the compressor unit and leave the other end open Then switch on the compressor unit and blow the tube dry The exterior of both the air tube and compressor unit should be cleaned...

Page 8: ...l socket Make sure that the ends of the air tube are fitted tightly onto the compressor unit and onto the medication tank Check whether the medication tank is empty If so and if necessary refill the medication tank Is the vaporiser head installed correctly in the medication tank If not install it and press it down into the jet nozzle Check whether the air tube is twisted If so untwist Check whethe...

Page 9: ...H Storage conditions Temperature 25 C 70 C Humidity 10 95 RH Supply voltage AC 230 V 50 Hz Power consumption 85 W Electrical safety Fuse T2 5AL250V UR Rit Class 2 Type B Contents of the package Compressor unit nebuliser kit JetAIRplus air tube flexible child mask flexible adult mask carrying case 5 spare filters instruction manual guarantee card value measured with 0 9 NaCl solution Please note th...

Reviews: