Removing condensation from the air tube
If there is moisture or liquid remaining in the air tube, be sure to follow the procedure
below to remove any moisture from within the air tube.
1) Make sure that the air tube is connected to the air connector on the compressor.
2) Remove the air tube from the nebulizer kit.
3) Turn on the compressor and pump air through the air tube to expel the moisture.
Changing the air filter
If the air filter has been used for more than 60 days, replace it with a new one.
1.
Pull the air filter cover off the compressor.
2.
Change the air filter.
3.
Put the air filter cover back in place.
Make sure it is attached properly.
Notes:
• Use only OMRON air filters designed for this device. Do not operate without air filter.
• Do not wash or clean the air filter. If the air filter becomes wet, replace it. Damp air filters
can cause blockages.
• There is no front or back orientation for the air filters.
• Check that air filters are clean and free of dust before inserting them.
Summary of Contents for COMP A-I-R C21 basic
Page 23: ...RU NE C803 E_A_M09_140715 pdf C21 basic NE C803 RU 2020 03 02...
Page 24: ...24 OMRON C21 basic 24 25 26 29 30 30 31 35 37 37 38 39...
Page 25: ...25 RU C21 basic 2 2 7 5 23 C 5 1 1 1 100 1 60 1...
Page 26: ...26 OMRON...
Page 27: ...27 RU 60 ee 45 10...
Page 28: ...28 30 OMRON HEALTHCARE 40 C...
Page 29: ...29 RU HHP CM01 x 1 x 1 x 1 x 5 100 x 1 x 1 x 1 x 1 x 1 x 1...
Page 31: ...31 RU 1 2 3 4...
Page 32: ...32 5 6 1 2...
Page 33: ...33 RU 7 8...
Page 34: ...34 9 45 10 11 45 45...
Page 35: ...35 RU 36...
Page 36: ...36 o X 100 A o o o o o X X o o o o o X X 0 1 15 B o o o o X X X A B 15 20...
Page 37: ...37 RU 1 2 3 60 1 2 3 OMRON...
Page 38: ...38 30 45 OMRON...
Page 40: ...40 IP IEC 60529 12 5 OMRON HEALTHCARE EUROPE URL www omron healthcare com BF II IP XX 60529...
Page 46: ...46 2 1 7 2790 10...
Page 47: ...47 RU...