ERROR INDICATORS
SYMBOL
CAUSE
CORRECTION
Monitor could not
detect pulse wave.
Take another measurement and
remain still until the measurement is
complete.
Cuff under-in
fl
ated.
Cuff not applied
correctly.
Remove the arm cuff.
Read “Taking a Measurement”.
Wait 2-3 minutes.
Take another measurement.
Cuff over-in
fl
ated.
Device error.
Contact Customer Service.
Movement during
measurement.
Keep still and take another
measurement.
Irregular or weak
pulses are detected.
Remove the arm cuff. Wait 2-3
minutes and then take another
measurement. Repeat the steps in
section “Applying the Arm Cuff”.
If this error continues to appear,
contact your doctor.
Batteries are
exhausted.
Replace the four batteries.
Refer to “Battery Installation”.