49
ImpOrTANT INFOrmATION rEgArdINg ELECTrO mAgNETIC COmpATIBILITy (EmC)
Recommended separation distance between portable and mobile RF communications equipment and the PM601
The PM601 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customers or the users of
the PM601 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the PM601 as recommended below, according to the maximum output power of the communications equipment.
Output Power of
Transmitter in Watt
Separation distance according to frequency of transmitter in meter
150 kHz to 80 MHz
d = 1.16 d = 0.58
80 MHz to 800 MHz
d = 0.35
800 MHz to 2.7 GHz
d = 0.7
0.01
0.12
0.06
0.04
0.07
0.1
0.37
0.18
0.11
0.22
1
1.16
0.58
0.35
0.7
10
3.67
1.8
1.1
2.2
100
11.6
5.8
3.5
7.0
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
Note:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies
Note:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects, and people.