47
ImpOrTANT INFOrmATION rEgArdINg ELECTrO mAgNETIC COmpATIBILITy (EmC)
Electromagnetic immunity IEC60601-1-2
Immunity test
IEC 60601 Test level
Compliance level
Electromagnetic environment –guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V rms
150 kHz to 80 MHz
6 V rms
The ISM bands and the amateur radio
bands between 150 kHz to 80 MHz
10 V/m
80 MHz to 2.7 GHz
IEC 60601-1-2:2014 Table 9
(Up to 28 V/m at certain
frequencies)
3 V rms
150 kHz to 80 MHz
6 V rms
The ISM bands and the amateur radio
bands between 150 kHz to 80 MHz
10 V/m
80 MHz to 2.7 GHz
IEC 60601-1-2:2014 Table 9
(Up to 28 V/m at certain
frequencies)
Portable and mobile RF communications equipment
should be used no closer to any part of the PM601
including cables, than the recommended separation
distance calculated from the equation appropriate to
the frequency of the transmitter.
Recommended separation distance
d
= 1.16
150 kHz to 80 MHz
d
= 0.58
150 kHz to 80 MHz
(The ISM bands and the amateur radio bands)
d
= 0.35
80 MHz to 800 MHz
d
= 0.7 800 MHz to 2.7 GHz
Summary of Contents for Avail PM601