13
O
MNIP
O
D
®
I
N
STR
UC
TION
S
If the PDM screen times out during the process, press and hold the
Home/Power
button to continue.
3. Press start
3.1
+
The Pod automatically inserts
the cannula and delivers a
prime bolus to fill the cannula
with insulin. It takes a few
seconds to complete this
process. Release the skin
after the cannula inserts.
3.2
+
Once complete, the PDM
indicates that the Pod is active
and asks you to check the
infusion site.
+
Look through the Pod’s viewing
window, if properly inserted,
press yes. Press No if you see a
problem with the cannula.
3.3
+
The PDM will generate
an automatic reminder to
check your blood glucose
1.5 hours after each Pod
change.
2.4
+
Carefully remove white paper
backing from the adhesive,
ensuring the adhesive is clean
and intact.
3:00p 5/15
Remove pod’s adhesive
backing
If pod is wet or dirty, or
adhesive is folded, press
“Discard.”
If pod OK, apply to site.
Back
Next
Discard
3:00p 5/15
Press “Start” to insert
cannula and begin basal
delivery.
Start
3:00p 5/15
Pod is active.
“basal 1” has been
programmed.
Check infusion site and
cannula.
Is cannula properly
inserted?
No
Yes
3:00p 5/15
Reminder:
Always check BG after a
pod change.
Check infusion site and
ensure cannula is
properly inserted.
OK
2.5
+
Remove and discard the
white paper backing from
the adhesive.
2.6
+
Apply the Pod to the selected site.
+
Run your finger around the
adhesive to secure it.
+
Press Next on the PDM.
+
To facilitate insertion, place one
hand over the Pod and make
a wide pinch around the skin
surrounding the viewing window;
this step is critical if the insertion
site does not have much
fatty tissue.
WARNING:
+ The PDM will generate an automatic reminder to check your blood glucose 1.5 hours after each Pod change. If the cannula is not
properly inserted, hyperglycemia may result. Verify there is no wetness or scent of insulin, which may indicate the cannula has dislodged.
+ NEVER inject insulin (or anything else) into the fill port while the Pod is on your body. Doing so may result in unintended or
interrupted insulin delivery.
+Verify cannula does not extend beyond adhesive backing once needle cap is removed.
OMNIPOD
®
INSTRUCTIONS |
Activate a new Pod