8.2 OPERATION (Continued)

 

Problem Possible 

Cause 

Action(s) 

Glass sample vessel 

cracked or  

shattered 

The processing tip is touching 

the side or bottom of the  

container 

1.  Do not allow the processing tip to 

touch the sample vessel. 

The processing tip is not  

immersed in enough fluid 

1.  Turn off the unit 
2.  Check the depth of the tip in the 

sample liquid.  

3.  Adjust the probe if necessary. 

The processing tip is fractured 

or damaged causing it to  

resonate at an incorrect  

frequency. 

1.  Turn off the unit 
2.  Remove the damaged processing tip. 
3.  Attached a new washer and tip. 

Loud piercing 

sound or  

inconsistent sound 

coming from the tip 

SECTION 8. TROUBLESHOOTING 

Summary of Contents for Sonic Ruptor 400

Page 1: ...ULTRASONIC HOMOGENIZER USER MANUAL OMNI SONIC RUPTOR 400 Rev A...

Page 2: ...Ruptor 400 For Use Section 4 Sonic Ruptor 400 Operation 4 1 Control Panel 4 2 Start Up of the Equipment 4 3 Power Output 4 4 Timer Settings 4 5 Pulser Settings 4 6 Start Reset Section 5 Operation 5 1...

Page 3: ...For the balance of the period parts will be provided but service will be charged at established labor rates This warranty does not include normal wear from use it does not apply to any instrument or...

Page 4: ...the instrument from the main power supply remove the power cord from the AC power input on the rear panel and clean the equipment and the surrounding area DO NOT reconnect the equipment until it has...

Page 5: ...1 Place the shipping box on a level surface 2 Remove the foam from around the control unit and accessories 3 Place the Sonic Ruptor on a clean horizontal and stable surface 4 Do not block the cooling...

Page 6: ...N 5 32 Stepped Micro Processing Tip Very High Intensity Diameter 5 32 3 8mm Length 10 1 25 6cm 250 L 10mL OR T 156 3 8 Intermediate Processing Tip High Intensity Diameter 3 8 9 5mm Length 8 6 21 8cm...

Page 7: ...d up to 8 micro tubes PRODUCT DESCRIPTION ORDER 250mL Cup Tip OR C 250 Microtube Tray for OR C 250 OR C 250 T PRODUCT DESCRIPTION ORDER Continuous Flow Chamber OR F 01 Continuous Flow Chamber The Cont...

Page 8: ...apors and water CAUTION Always place a washer between the Tip and the BNC Transducer Attaching the Processing Tip Sold Separately Remove the protective plastic caps from the Processing Tip and the Tra...

Page 9: ...power entry module located on the rear panel of the control unit Plug the power cord into a grounded outlet of 115 volts 60 cycles or 230 volts 50 cycles AC Check the Serial Plate and power entry modu...

Page 10: ...t of ultrasonic power being generated by the Processing Tip The power output is adjusted by turning the Power knob to the right or left The ten segment LED Power Output Meter reflects the level of the...

Page 11: ...ate of one pulse per second The pulse duration can be adjusted from 10 to 90 enabling a solution to be processed at full power range while limiting temperature The pulse mode is especially valuable fo...

Page 12: ...vertor If the Timer is used the processing will automatically stop when the Timer is through Push the Reset button to terminate the processing at any time Transducer The Transducer assembly may be han...

Page 13: ...ove the threshold of cavitation audible sounds are emitted from the vessel containing the liquid These sounds can be irritating to the operator and care should be taken to avoid prolonged exposure For...

Page 14: ...the fan ALWAYS disconnect the power cord before cleaning DO NOT use any type of scrapers DO NOT use caustic solvents or acetone Harsh solvents can damage the finish of the housing Example of Decontami...

Page 15: ...certificate and attach to the instrument or parts being returned If no Decontamination Certificate is attached and a potential radioactive or biological hazard is detected or suspected by Omni Interna...

Page 16: ...cer 3 Replace washer The probe is not properly secured 1 Turn off the unit Use the wrenches to tighten the processing tip on the transducer Burning smell from transducer 1 Turn off the control unit 2...

Page 17: ...sel The processing tip is not immersed in enough fluid 1 Turn off the unit 2 Check the depth of the tip in the sample liquid 3 Adjust the probe if necessary The processing tip is fractured or damaged...

Page 18: ...earch Please visit the web address below to register your new Sonic Ruptor You will also be asked to provide us with information about your research In return we ll send you a Starbucks gift card Regi...

Page 19: ..._________________________________________ ___________________________________________________________ ___________________________________________________________ ______________________________________...

Page 20: ...20 935 C Cobb Place Blvd Toll Free 1 800 776 4431 Kennesaw GA 30144 Phone 770 421 0058 Patents Pending P N 03 254 www omni inc com...

Reviews: