Integration into systems
The
MPD 800
acquisition unit,
CPL1
,
CPL2
and the
V-to-AC-adapter
are designed as a system. It is
recommended to use them as described in the respective “Designated use” (see 2.2) and “Application”
(see 2.3) chapters. If one or more of the
MPD 800
system components need to be integrated into a
system, the following additional precautions must be fulfilled:
► If connected to an existing coupling capacitor, locate external quadripoles
CPL1
and
CPL2
in close
proximity of the bottom end connection of the coupling capacitor with a distance of less than one
meter.
► If a ground-referenced measurement device which is not part of the
MPD 800
system is used
together with the
MPD 800
system, use
CPL2
due to special grounding restrictions. The maximum
ground-referenced voltage at the input of the low arm capacitor must not exceed 140 V
RMS
under any
measurement condition. Therefore, perform additional precautions to achieve the desired safety
requirements.
► Do not use neither
CPL1
nor
CPL2
in an electrical path where a failure of
CPL1
or
CPL2
(for example,
short or open circuits of measurement inputs) causes:
• a voltage regulation feedback loop to become dysfunctional.
• an over-current protection to become dysfunctional.
• an indicator for hazardous live electrical quantities (IEC 61010-1, VDE 0105-100) to become
dysfunctional.
► Ground
CPL1
or
CPL2
using the shortest path with the lowest achievable inductance. For this
purpose, use the common grounding circuit of the test voltage source, test object and coupling
capacitor.
► The
MPD 800
acquisition unit and
RBP1
are inherently grounded through the connection cables to
CPL1
or
CPL2
. Therefore, make sure that the device housings is electrically insulated from ground
potential.
► Place
MCU2
in close proximity to the PC and outside of the high voltage or test area (EN 50191).
Make sure that the computer is grounded through protective earth, for example, grounded USB hub.
► It is not recommended to place
CPL1
or
CPL2
outside of the high-voltage or test area (EN 50191). If
CPL1
or
CPL2
is placed outside these areas, perform additional precautions in order to achieve an
equivalent safety level according to safety standards (see 1.2.1 "Safety standards" on page 6).
DANGER
Death or severe injury caused by high voltage or current
► Always observe the five safety rules (see 1.2.2 "Safety rules" on page 6) before
connecting or integrating
CPL1
or
CPL2
into a system.
► Provide protection against live parts in order to avoid accidental contacts.
► Connect the
AC
output of
CPL1
or
CPL2
to the
MPD 800
acquisition unit or short-
circuit the
AC
output because the
AC
output can have voltages higher than 30 V. The
AC
output is limited by a protection circuit.