MPD 800 User Manual
18
OMICRON
2.2.2
CPL1
and
CPL2
external quadripole
Designated use
Figure 2-5:
CPL1
and
CPL2
external quadripole
The
CPL1
and
CPL2
external quadripoles are designed to connect the
MPD 800
acquisition unit to a
capacitance (coupling capacitor or test object), which drives a reactive current larger than the current the
MPD 800
acquisition unit can handle directly.
CPL2
allows furthermore to operate third party measuring equipment using the same coupling capacitor
or test object as the
MPD 800
system. Additional precautions must to be fulfilled to achieve the desired
safety requirements.
DANGER
Death or severe injury caused by high voltage or current
► Always observe the five safety rules (see 1.2.2 "Safety rules" on page 6) before
connecting or integrating
CPL1
or
CPL2
into a system.
DANGER
Death or severe injury caused by high voltage or current
CPL1
or
CPL2
and its accessories are not suitable for reliably checking if a system is
electrically live or energized.
► Do not use
CPL1
or
CPL2
for safety relevant procedures.