OMICRON
19
MPD 800 measurement system
Front panel connections
Figure 2-6:
CPL1
and
CPL2
front panel
The BNC outputs
PD
and
AC
are connected by short, screened cables (< 20 cm) to the BNC inputs at
the
MPD 800
acquisition unit accordingly (see "Rear panel connections" on page 14).
Rear panel connections
CPL1
Figure 2-7:
CPL1
rear panel
The GND or GND wing nut is connected to the ground with a short connection.
► Do not ground the
MPD 800
acquisition unit or its accessories in this case.
The
IN
input is connected to the bottom end connection of the coupling capacitor or test object.
WARNING
Death or severe injury caused by high voltage or current possible
When the test object is energized flashovers or electrical breakthroughs may occur due
to failure of test object. The
MPD 800
acquisition unit and
CPL1
or
CPL2
are not
designed to withstand power frequency fault currents.
► Always observe the minimum safety distance and use personal safety equipment.
WARNING
Death or severe injury caused by high voltage or current possible
An open
AC
output may carry hazardous voltage.
► Terminate the
AC
output using a short-circuit cap.