Increasing the Output Power, Operating Modes
93
7.2.2
1 x 64 A High Burden and High Current Mode (L-L)
1 x 0 ... 64 A (±90 A
DC
), max. 70 V
peak
, 1 x 1740 VA at 50 A
The currents 1 and 2 of each group are phase-opposite. In addition, the
groups A and B are connected in parallel.
Observe the safety instructions given in Section 7.1 on page 91 when using
this operating mode.
Figure 7-2:
Single-phase operation,
1 x 64 A high burden and
high current mode
Refer to the output curves shown in the figures 6-1 through 6-5 in section
6.3.2, "Current Outputs" on page 43.
Load
1’
N’
Summary of Contents for CMC 356
Page 1: ...CMC 356 User Manual...
Page 38: ...CMC 356 Reference Manual 38...
Page 100: ...CMC 356 Reference Manual 100...
Page 104: ...CMC 356 Reference Manual 104...
Page 132: ...CMC 356 Reference Manual 132...
Page 136: ...The OMICRON Bootloader Software Copyright Notices 136...