7.1.2
Current Amplifier CMA 156
The 6-phase current amplifier
CMA 156
serves as an additional device to the
CMC 156
test set, for tests that require:
• higher currents and/or higher power in the current path than is delivered by
the
CMC 156
• up to nine independent current phases.
The six current phases are arranged in two isolated groups and can be used
independently from the phases of the
CMC 156
(e.g. for differential protection
testing). By connecting the current phases in series, a power of up to 280 VA
can be output, which allows the testing of a wide range of electromechanical
relays.
The
CMA 156
is connected via the interface Gen. out 7-12 on the back of the
CMC 156
test set:
Figure 23
Connection of the
CMA 156
current amplifier to the
CMC 156
test
set
CMA 156
current amplifier
CMC 156
test set
Ampl. in
Gen. out
7-12