Connectors and Interfaces
27
4.2 Connections on the Back Panel
Figure 9
Rear view
CMC 156
4.2.1
Gen. out 7-12
At the socket Gen. out 7-12 two completely independent generator triples are
supplied. These signals can be used to control an external amplifier or to be
used directly as low level signal outputs.
For connection to this socket use the cable which is included in the scope of
delivery of the external amplifier (connection cable
CMC 156
- amplifier).
The outputs are isolated from mains and the power outputs by reinforced
isolation (SELV interface) and supply calibrated signals in the range 0 to 5V
rms
(7.07V
pk
).
The selection of the amplifiers as well as the adaptation of the scaling for the
corresponding amplifier to be connected takes place in the software.
Corresponding notes are found in chapter 3 of the CMC Software manual.
Host interface
Fan
Fan
Fuse
T6.3AH
Gen. out 7-12
Ext. Interface
Power-
supply