Safety Instructions

2

Safety Instructions

Before using C-Probe 1, please read the following safety instructions 
carefully.

Without a detailed knowledge about the safety instructions, C-Probe 1 may 
not be set into operation!

C-Probe 1 may only be operated by authorized personnel.

2.1 

Principle Rules for Usage

• C-Probe 1 should only be used when in a technically sound condition. Its 

use should be in accordance with the regulations, in a safe manner, and 
mindful of the dangers. In particular, avoid disruptions that could in turn 
affect safety.

• C-Probe 1 is exclusively intended for the measurement of DC and AC 

currents within the ranges specified in this operation manual (refer to 3.3 
“Input and Output Quantities” on page 10)
. Any other use is deemed as 
not being according to the regulations. The manufacturer/distributor are 
not liable for damage resulting from unintended usage. The user alone 
assumes all responsibility and risk.

• Adhering to the instructions provided in this operations manual is also 

considered part of being accordance with the regulations.

2.2 Safety 

Operation

• Never use the current clamp at substations which carry a potential higher 

than 600 V with respect to earth!

• Never put the current clamp around any live wires if it is not connected to 

a measurement device!

• When grasping cables or current busbars with the current clamp, make 

sure that they are well-centered and perpendicular to the clamp!

• Keep the enclosed wire as far as possible from adjacent wires to prevent 

distortions due to magnetical interference fields!

• When replacing the battery, the current clamp power switch has to be in 

the "OFF" position and the clamp must be separated from busbar and 
the measurement device!

• Do not use the current clamp if it is damaged or you detect indications of 

a possible defect! This may be hazardous to life.

Summary of Contents for C-Probe 1

Page 1: ...27 AC DC Current Clamp C Probe 1 User Manual...

Page 2: ...a embodied in this manual represent the technical status at the time of writing and are subject to change without prior notice We have done our best to ensure that the information given in this manual...

Page 3: ...ut Quantities 10 3 4 Typical Noise Levels at Output RMS 11 3 5 Influencing Factors 11 3 6 Voltage Supply 11 3 7 Environmental Conditions 12 3 8 Safety 12 3 8 1 Electrical Safety 12 3 8 2 Electromagnet...

Page 4: ...AC DC Current Clamp C Probe 1 4 6 Maintenance and Service 23 6 1 Storage 23 6 2 Cleaning 23 6 3 Service and Calibration 23 7 Appendix Diagrams 25 7 1 Ordering information 25 7 2 Diagrams 25 Index 31...

Page 5: ...eople working with C Probe 1 In addition to the operational procedures described in this user manual follow the specific safety regulations for accident prevention applicable in the country and at the...

Page 6: ...AC DC Current Clamp C Probe 1 6...

Page 7: ...ibutor are not liable for damage resulting from unintended usage The user alone assumes all responsibility and risk Adhering to the instructions provided in this operations manual is also considered p...

Page 8: ...to faulty material or manufacturing defects Under no circumstances can the claim under warranty exceed the amount listed in the bill Therefore the demands are limited to the repair of the defective de...

Page 9: ...9 Technical Data 3 Technical Data 3 1 Dimensions Figure 3 1 Dimensions of C Probe 1...

Page 10: ...the core and hence to faulty measurements This does not affect safety 0 10 A AC DC 0 40 A AC DC 40 60 A AC DC 60 80 A AC DC Accuracy 10 100Hz 1 of meas value 5 mA 2 of meas value 30 mA 0 8 of meas val...

Page 11: ...r it was switched on 3 4 Typical Noise Levels at Output RMS Table 3 2 Typical noise levels 3 5 Influencing Factors Position of wire in the clamp Max 0 5 of displayed value Wire from stop to stop Tempe...

Page 12: ...50 60 Hz between hand held area at normal usage and primary winding secondary output 3250 V 50 60 Hz between primary winding and secondary output Maximum voltage according to IEC 1010 1 600 V in inst...

Page 13: ...class 2 without interference 2 kV 1 kA class 3 with low interference according to IEC 801 5 3 8 3 Mechanical Safety Saftey class IP 20 according to IEC 529 Shock resistance up to 100 g half period of...

Page 14: ...AC DC Current Clamp C Probe 1 14...

Page 15: ...ith the circuit to be measured The current measurement range reaches from 50 mA up to 80 A AC and DC The current clamp contains two measurement ranges 10 mV A and 100 mV A as well as two control lamps...

Page 16: ...gend 1 Slide switch with 3 positons OFF 10 mV A 100 mV A 2 Battery condition indicator 3 Zero adjustment of the current clamp 4 Battery compartment 5 Connecting cable 2 meter with banana plugs 6 Contr...

Page 17: ...al error of the current clamp occured 3 Zero adjustment of the current clamp By means of this thumbwheel the output voltage of the current clamp can be adjusted to zero This is necessary to compensate...

Page 18: ...ort current peaks exceed the measurement range the measurement is not valid due to other reasons 7 Direction arrow The arrow symbol shows the reference flow direction of the DC current to be measured...

Page 19: ...nsates thermal effects remanent magnetic fields or the geomagnetic field Figure 5 1 Zero adjustment 1 Connect C Probe 1 to a DC voltmeter e g to the Analog DC Input of the CMC 256 test set 2 Select a...

Page 20: ...surement range 5 2 Current Measurement Figure 5 2 Current measurement 1 Set the slide switch of C Probe 1 to the desired measurement range of 100 mV A or 10 mV A The control indicator ON lights up 2 P...

Page 21: ...ry as described in the following section 5 4 Battery Replacement Before replacing the battery please ensure the following The current clamp may not grasp any wire and must be disconnected from any mea...

Page 22: ...AC DC Current Clamp C Probe 1 22...

Page 23: ...ibration Should there be a reason for complaint please contact our service department or local subsidiary and give them a detailed report of the reason for the complaint The addresses are listed in se...

Page 24: ...AC DC Current Clamp C Probe 1 24...

Page 25: ...10 A AC DC The given accuracy data refer to the measured value An additional measurement error of 5 mA must be taken into account Order description Order number C Probe 1 AC DC current clamp Supplied...

Page 26: ...measured value An additional measurement error of 30 mA must be taken into account Maximaler Phasenfehler im Me bereich 100 mV A 0 10A 35 30 25 20 15 10 5 0 5 10 100 1000 10000 Frequency Hz Max error...

Page 27: ...value An additional measurement error of 30 mA must be taken into account Maximaler Phasenfehler im Me bereich 10 mV A 0 40A 35 30 25 20 15 10 5 0 5 10 100 1000 10000 Frequency Hz Max error degree 0...

Page 28: ...m phase error at measurement range 10 mV A for measurement currents of 40 80 A AC Maximaler Amplitudenfehler im Me bereich 10 mV A 60 80A 14 12 10 8 6 4 2 0 2 1 10 100 1000 10000 Frequency Hz Max erro...

Page 29: ...Ltd Hong Kong Phone 852 2634 0377 E Mail support asia omicron at Web www omicron at North and South America OMICRON electronics Corp USA Phone 1 713 830 4660 or 1 800 OMICRON E Mail techsupport omicr...

Page 30: ...Contact Information Technical Support 30...

Page 31: ...ntact information OMICRON address 29 C Probe 1 calibration 23 cleaning 23 current measurement 20 current measurement range 15 dimensions 10 environmental conditions 12 functional description 15 genera...

Page 32: ...C 26 amplitude error at measurement range 10 mV A for measurement currents of 40 60 A AC DC 27 amplitude error at measurement range 10 mV A for measurement currents of 60 80 A AC DC 28 amplitude error...

Page 33: ...nge 15 measurement range overflow 18 measurement range overflow indicator 18 measurement range current 15 mechanical safety 13 N noise level at output 11 O opening of clamp jaws 10 operating elements...

Page 34: ...34 T technical data 9 Technical Support 29 temperature variation 11 transformer current potential 15 V voltage on live wire 10 voltage supply 11 W warranty 8 weight 10 wire diameter 10 wire position...

Reviews: