![Omicron C-Probe 1 User Manual Download Page 12](http://html1.mh-extra.com/html/omicron/c-probe-1/c-probe-1_user-manual_3977006012.webp)
AC/DC Current Clamp C-Probe 1
12
3.7 Environmental
Conditions
Figure 3-2:
Environmental conditions
for C-Probe 1
3.8 Safety
3.8.1 Electrical
Safety
• Device with double isolation in hand-held area at normal usage, and
single isolation between primary winding and secondary output.
• High voltage test (according to IEC 1010-1)
- 5550 V (50/60 Hz) between hand-held area at normal usage and
primary winding / secondary output.
- 3250 V (50/60 Hz) between primary winding and secondary output.
• Maximum voltage (according to IEC 1010-1)
- 600 V in installation category III, degree of pollution 2.
- 300 V in installation category IV, degree of pollution 2.
Environm e ntal conditions
0
10
20
30
40
50
60
70
80
90
100
-50
0
50
100
Te m pe rature / °C
R
elat
ive humidit
y / %
Environmental conditions
0
20
40
60
80
100
-50
0
50
100
Temperature / °C
Relative humidity / %
Reference area
Operation
Storage