Calibration Instructions
3.1 Equipment Required
• Precision mV source, with 0.001 mV resolution and ±0.002 mV
accuracy or
• Precision DVM with ±0.002 mV accuracy and an adjustable mV
source with 0.001 mV resolution
• OMEGA TRC III Ice Point Reference (or stable ice bath)
• Temperature Reference Probe (OMEGA P/N: TRP-(*)
*Thermocouple Type J, K, T, E