19. HARDWARE CHECK
19.7 Communications IF Check
2
Check Flow Chart
Operation screen
[Engineering mode selection]
( + )
3 sec. or more
[Selecting check]
→
[Displaying check 1]
[Displaying check 2]
[Selecting communications]
[Communications check]
Check ends
Press or to switch
Check 1 and 2.
(1)
: Alarm output check
(2)
: Remote contacts input check
(3)
: Communications interface check
1. Communications type
Communications type installed is
displayed.
•
RS232C
…
•
RS422A
…
•
RS485
…
2. Communications status display
•
Normal condition:
Displays one digit by one digit as
0
→
1
→
2 …9, and after 9,
repeat from 0 again.
•
Abnormal condition:
One of 0 to 9 blinks. In this case,
communications circuit is
abnormal.
2 sec. or more