7. SET UP MODE
7.10 Turning Alarm Printing On/Off
7.10 Turning Alarm Printing On/Off
When alarms occur, the channel number, level, type of alarm and ON/OFF time can be
printed on the right side of the chart. It is selectable to get the ON/OFF time of alarms
(ON1), the ON time only (ON2) or to get no printout (OFF). Note that alarms will not
be printed when the chart speed exceeds 1800 mm/h (Pen model) or 120 mm/h (Dot-
printing model).
The initial value is ‘ON1’, and this setting is common for all channels.
SETTING PROCEDURE:
Enter the SET UP mode by turning ‘ON’ the power while pressing and holding the [CH
UP]
(
ENT)
key until the SET UP display appears.
Use the [
] key to select the display [
]. Press the [ENT] key.
Use the [
] key to select the display [
]. Press the [ENT] key.
Select ‘ON1’ (ON/OFF time of alarms will be printed), ‘ON2’ (ON time only will be
printed) or ‘OFF’ (no alarm will be printed). Press the [ENT] key.
The display [
] appears.
Before leaving the SET UP mode, you have to store your new settings. Use the [
] key
to select the [
] display, and press the [ENT] key.
Select [
] to keep your new settings or [
] to abort the settings, and
press the [ENT] key. After a few seconds, the OPERATION mode will appear.