6 - 6
Digital Input Setting (DI Setting)
If a Digital Input (contact input/voltage level input) is being used, the setting can be
done as follows:
Key operation
Indication
Description
Press the [DISP] key for three seconds to enter the SET mode.
Select the [
] display by using the [
] key.
Then press the [ENT] key.
Use the [
] key to select the desired channel.
Press the [ENT] key.
Use the [
] key to select [
].
Press the [ENT] key.
There are two types, namely LEVEL and CONTACT.
LEVL: (Level)
Records ON/OFF status of a voltage input
(0 to 2.4V is OFF (0), 2.4V and higher is
ON (1)).
CONT: (Contact) Records ON/OFF status of a contact (close
is ON (1), open is OFF (0)).
Select the desired type using the [
] key and press the [ENT]
key.
[
] is displayed and the setting is completed.
Press the [ENT] key to set another channel; or press the
[DISP] key for three seconds to return to the OPERATION
mode.
NOTE
• The above mentioned selections will result in a recording trace from 0mm to
100mm between the contact statuses OFF and ON, which is from the extreme
left side of the chart to the extreme right side.
• If you press the [ESC] key in the middle of setting, the indication returns to the
[
] and aiready set data will be aborted.
DISP
ESC
3 sec
DISP
ESC
3 sec
CH UP
CH UP
CH UP
CH UP
CH UP