3.700
mA AOLC
modify
parameters
S
Current parameters
enter settings
Z
Ran 3.500~3.800
:
Ran 3.500~3.800
:
3.700
mA AOLC
Ran 3.500~3.800
:
3.701
mA AOLC
confirm and
enter next menu
M
Display selected parameters
3.500-3.800mA
Display modified
parameters
confirm and enter next menu
Z
After above operation, press M get into DisMod-LRV-Unit parameters setting,and press M again get into hidden parameters setting
0
FixTOut
0
0
1
0
1
FixTOut
enter settings
Z
Measuring Output
Fix COut
Measuring Output
Fix COut
Output 3.800mA
Fix COut
Both off
Current parameters
Display selected parameters
confirm the output
Z
Measuring interface
0.010
kPa
0
1
2
3
4
5
6
7
Measuring Output
Output 3.800mA
Output 4.000mA
Output 8.000mA
Output 12.000mA
Output 16.000mA
Output 20.000mA
Output 20.800mA
0
1
2
3
Both Off!
Fist On!
Second On!
Both On!
Fix TOut
Current parameters
Display selected parameters
Display modified
parameters
Display modified
parameters
23.000
23.000
23.001
confirm and enter next menu
Z
enter settings
Z
Ran 20.800~24.000
:
mA AOHC
Display selected parameters
mA AOHC
Ran 20.800~24.000
:
Ran 20.800~24.000
:
mA AOHC
20.800-24.000mA
Current parameters
Display modified
parameters
Note 1: By default, the system shows optimal precision according to measuring range. Eg: While the measuring range of product is 0-10kPa,
measuring value shows 0.000-10.000; while the measuring range of product is 0-100kPa, the measuring value shows 0.00-100.00. If DotPos
parameters value and measuring range do not match, the system will automatically ignore and then process by default.
Eg: While the measuring range of product is 0-10kPa, if operator makes DotPos=5, the system will regard it as ineffectiveness and the measuring
value will still shows 0.000-10.000 by default; while the measuring range of product is 0-100kPa, if operator makes DotPos=4 or 5, the system will
regard it as ineffectiveness and the measuring value will still shows 0.00-100.00 by default.
Note 2:Use FixCOut parameters to control fixed current ,to confirm whether current output is ok. Changing fixed current value by press S and confirm
by Z.
Note 3:Use FixTOut parameters to control 2 way PNP by mannual operation, to confirm whether PNP output is ok. Changing fixed current value by
press S and confirm by Z.