Power

Signal

blue/3

+

SW1

S+

-

SW2

brown/1

white/2

black/4

gray/5

Label 

1

5

2

3

4

1

2

4

6

3

R2

R1

4/black wire is the common port              

Connect the positive power supply (P+) to the terminal 1/brown 
wire of pressure switch;  

Connect the negative signal module (S-) to the negative power 
supply(P-);

Connect the negative signal module (S-) to the negative power 
supply(P-) and then connect to the terminal 4/black wire of 
pressure switch;

Connect the positive signal module (S+) to the terminal 3/blue wire 
of pressure switch;

Connect the first way transistor to the terminal 2/white wire of  
pressure switch;

Connect the second way transistor to the terminal 5/gray wire of  
pressure switch.

1

2

3

4

5

6

A zero point adjustment should be made after installation to counter 
any effects the mounting position may have on the zero setting.
When performing the adjustment, the vessel should be completely 
empty, with no pressure or media on the measuring diaphragm and the 
vessel exposed to atmospheric air. With unit powered, short the key-z 
wire (blue) to the + power lead. hold for 5 seconds, when removed the 
zero point will be adjusted.Set the zero point three weeks after 
installation to ensure accuracy, and reset zero point annually.

Standard current signal transistor output: 12-30VDC,

Transistor output: 12-30VDC.

A cable with a shielded twisted-pair signal is recommended to 
avoid ground loop.

The built-in transient resistance module is only effective if properly 
grounded. The metal body and internal grounding terminals should 
be directly grounded.

To safeguard against liquid media potentially traveling along the cable 
into the terminal box, the cable should be configured in a U-shape as 
shown, with the bottom of the “U” lower than the bottom of the pressure 
switch.

Do not store outside.

Keep dry and dust-free.

Do not expose to the corrosive medium.

Avoid solar radiation.

Avoid mechanical shock and vibration.

Storage temperature: -40-185 (-40-85℃).

Maximum relative humidity: 95%.

EMC equipment instructions 2014/30/EU.

CE mark suggests the instruments are in line with EU standards

Users need to ensure the whole equipment conform to all the 
applicable standards.

Use a cleaner that is compatible with pressure switch materials

Clean carefully to ensure that the pressure port isn't damaged (e.g. 
with a sharp object)

Do not directly spray electrical connection or gauge vent with water 
or cleaner.

Please observe the following when cleaning:

A zero point adjustment should be made after installation to 
counter any effects the mounting position may have on the zero 
setting.

When performing the adjustment, the vessel should be completely 
empty, with no pressure or media on the measuring diaphragm and 
the vessel exposed to atmospheric air. 

With unit powered, short the key-z wire (blue) to the + power lead. 
hold for 5 seconds, when removed the zero point will be adjusted. 

Set the zero point three weeks after installation to ensure 
accuracy, and reset zero point annually.

Set PV=0 each year.

Summary of Contents for PSW2100

Page 1: ...INSTRUCTION SHEET For latest product manuals omegamanual info Shop online at omega com e mail info omega com PS 2100 M 5762 0320 W ...

Page 2: ...n be installed with adapters on the pipe directly Can be used for liquid level measurement in open container Media compatibility should be considered Two pressure switches can be used to create a differential measurement system commonly used in the filter control or closed container level measurement Light weight pressure can be mounted directly on the pressure leading tube When the maximum torque...

Page 3: ...lt in transient resistance module is only effective if properly grounded The metal body and internal grounding terminals should be directly grounded To safeguard against liquid media potentially traveling along the cable into the terminal box the cable should be configured in a U shape as shown with the bottom of the U lower than the bottom of the pressure switch Do not store outside Keep dry and ...

Page 4: ...i Normal setting SP1 145Psi RP1 137 78Psi Mode1 1 SPDT1 1 RPDT1 1 Pressure rises to 145Psi delay 1s OUT1 active on pressure drops to 137 78Psi delay 1s OUT1 reset off Output alarm signal when pressure is lower than 145Psi Normal setting RP1 145Psi SP1 152 28Psi Mode1 2 SPDT1 1 RPDT1 1 Pressure drops to 145Pssi delay 1s OUT1 active on pressure rises to 152 28Psi delay 1s OUT1 reset off Low level al...

Page 5: ... Psi Press at the same time at least 3 seconds Z S 0 010 Rotating Success 010 0 Measuring interface Press at the same time at least 3 seconds S M Finished interface Measuring interface Press at least 3 seconds M Measuring interface with lock Successful interface Finished interface 0 010 Psi Press M 0 010 Psi Press at least 3 seconds S 100 00 M key for exit 0 010 Psi Measuring interface Measuring i...

Page 6: ... 10 01 Ran 99999 99999 confirm and enter next menu Z Ran 99999 99999 Ran 99999 99999 URV Ran 99999 99999 URV kPa kPa Mpa Unit Unit confirm and enter next menu M Display selected parameters Display selected parameters confirm and enter next menu Z 0 1 2 3 4 5 6 7 8 9 10 11 12 0 1 2 Pressure Unit Two rows display PV Unit Three rows display percentage pv unit Three rows display SP1 RP1 alternate disp...

Page 7: ...ode1 enter settings Z Current parameters 0 0 0 No Out 0 1 Hysteresis P SPn Ran 0 00 60 0S SPDT1 Ran 0 00 60 0S 0 0 SPDT1 Display selected parameters Ran 0 00 60 0S 0 1 SPDT1 confirm and enter next menu Z No Out Mode1 Mode1 Measuring interface Display selected parameters Display modified parameters 0 0 60 0S 0 1 2 3 4 No Out Hysteresis P SPn Hysteresis P RPn Window RPn P SPn Window P RPn P SPn SP1 ...

Page 8: ...isplay modified parameters Display modified parameters enter settings Z modify parameters S confirm and enter next menu Z 0 0 Ran 0 00 60 0S 0 0 SPDT2 Ran 0 00 60 0S SPDT2 Ran 0 00 60 0S 0 1 SPDT2 Display modified parameters Current parameters Display modified parameters enter settings Z modify parameters S SP2 enter settings Z modify parameters S Measuring interface 0 010 kPa Press Z Current para...

Page 9: ...20 800 24 000mA Current parameters Display modified parameters Note 1 By default the system shows optimal precision according to measuring range Eg While the measuring range of product is 0 10kPa measuring value shows 0 000 10 000 while the measuring range of product is 0 100kPa the measuring value shows 0 00 100 00 If DotPos parameters value and measuring range do not match the system will automa...

Page 10: ...8 ...

Reviews: