7. 4-20 mA Transmitter Option

When equipped with this option the gauge no longer operates from batteries 

but instead is “Loop Powered”. The loop connection is made to a 6 pin 
receptacle located at the rear of the unit (a mating plug is supplied). A voltage of 
between 9 and 32 VDC must be maintained at this connection (Pin A is “+” and 
Pin B is “-” see sketch) to insure proper operation. Completion of the earth or 
system ground (Pin F) is recommended for proper circuit protection.

Power supply voltage must be sufficient to maintain a minimum of 9 VDC 

at the gauge terminals after “dropping” voltage across RL at full scale current 
(20 mA). Example: If RL = 250 ohm then “drop” is 0.02 Amps X 250 ohm = 5 
volts. Therefore power supply minimum is 5 V + 9 V = 14 V.

RE-CALIBRATION: The procedure is the same as in step 4, except that 

there are 2 sets of zero and span adjustments. The front panel controls affect the 
display and the rear controls (remove “battery” door) affect the 4/20 mA signal.

A

B

C

D

– EXCITATION

+ EXCITATION

Gauge

Connector

Typical 4-20 mA Circuit

4-20 mA

CURRENT

LOOP

E

F

Controller

Process

L

R

+Vin

 

DPG5000 

M3366/0606

6

Summary of Contents for PRESSURE GAUGES DPG5000

Page 1: ...DPG5000 PRESSURE GAUGES M3366 0606 User s Guide http www omega com e mail info omega com...

Page 2: ...ine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Conductivity Eq...

Page 3: ...mming 4 Use of Auto Tare Feature 4 Battery Replacement 4 Re Calibration 5 Use Of Peak Hold Feature Optional 5 Analog Output Option 0 To 2 VDC 5 4 20 mA Transmitter Option 6 External Power 24 VDC and 0...

Page 4: ...GA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA either ver...

Page 5: ...kaging to prevent breakage in transit Warranty Returns For WARRANTY RETURNS please have the following information available BEFORE contacting OMEGA 1 Purchase Order number under which the product was...

Page 6: ...he Span should only be changed as part of the re calibration of a gauge with a known pressure source 2 Use of Auto Tare Feature When the DPG5000 is first turned on you will notice that the display fla...

Page 7: ...hold feature and causes an arrow to appear in the display In this mode only the highest reading sensed will be shown until a higher reading comes along This peak reading will be retained until the uni...

Page 8: ...rcuit protection Power supply voltage must be sufficient to maintain a minimum of 9 VDC at the gauge terminals after dropping voltage across RL at full scale current 20 mA Example If RL 250 ohm then d...

Page 9: ...T PIN D MINUS 24 VDC POWER IN PIN E NO CONNECTION PIN F CASE GROUND 9 External Power Without Other Options The external power input is accessed through a miniature Phone Jack mate supplied on the back...

Page 10: ...DPG5000 M3366 0606 Notes...

Page 11: ...irectives OMEGA will add the CE mark to every appropriate device upon certification The information contained in this document is believed to be correct but OMEGA Engineering Inc accepts no liability...

Page 12: ...44 e mail nl omega com Czech Republic Rud arm dy 1868 733 01 Karvin TEL 420 0 69 6311899 FAX 420 0 69 6311114 Toll Free in Czech Rep 0800 1 66342 e mail czech omega com France 9 rue Denis Papin 78190...

Reviews: