4. Re-Calibration

This procedure requires a known pressure source of at least ±0.1% accuracy in 

order to fully utilize the accuracy potential of the DPG5000 (If not available, the 
gauge can be returned to OMEGA for re-calibration.)

  Procedure:

A.  Ensure the Gauge is at 0 psig (or vacuum if absolute), and adjust the zero 

as per instructions in step 1.

B.  Apply full scale pressure to the pressure port and adjust the span (“S”) pot 

until the display reads the correct pressure.

C.  Re-check the zero and re-adjust the zero (“Z”) pot if required.
D.  Repeat steps B and C, until no further adjustment is required.

5. Use Of Peak-Hold Feature (Optional)

After initially turning the gage on (wait until the display stops flashing), a 

second push of the On/Off button energizes the peak-hold feature and causes 
an arrow to appear in the display. In this mode, only the highest reading sensed 
will be shown (until a higher reading comes along). This “peak” reading will be 
retained until the unit is reset by turning it off with a third push of the button. 
The sequence can then be re-initiated; one push for on in normal mode, two 
pushes for “peak-hold” mode and a third push for Off/Re-set.

6. Analog Output Option (0 To 2 VDC)

The 0–2 VDC analog output is accessed through a miniature “Phone Jack” 

(mate supplied) on the back of the unit (see External Power and 0-2 VDC 
Output Plug drawing on the last page). The “Tip” connection is “+ Output” 
and the center connection is “- Output”. DO NOT CONNECT TO THE CASE 
CONNECTION. This should only be used as a cable restraint. To minimize the 
effect on the battery life, the output should not be loaded with less than 100,000 
ohms.

0 to 2 VDC Table
 

 

0-2 VDC Output

 

Tip 

+ Output

 Center 

– Output

  Body 

No connection

 

 

M3366/0606 

DPG5000

5

Summary of Contents for PRESSURE GAUGES DPG5000

Page 1: ...DPG5000 PRESSURE GAUGES M3366 0606 User s Guide http www omega com e mail info omega com...

Page 2: ...ine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Conductivity Eq...

Page 3: ...mming 4 Use of Auto Tare Feature 4 Battery Replacement 4 Re Calibration 5 Use Of Peak Hold Feature Optional 5 Analog Output Option 0 To 2 VDC 5 4 20 mA Transmitter Option 6 External Power 24 VDC and 0...

Page 4: ...GA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA either ver...

Page 5: ...kaging to prevent breakage in transit Warranty Returns For WARRANTY RETURNS please have the following information available BEFORE contacting OMEGA 1 Purchase Order number under which the product was...

Page 6: ...he Span should only be changed as part of the re calibration of a gauge with a known pressure source 2 Use of Auto Tare Feature When the DPG5000 is first turned on you will notice that the display fla...

Page 7: ...hold feature and causes an arrow to appear in the display In this mode only the highest reading sensed will be shown until a higher reading comes along This peak reading will be retained until the uni...

Page 8: ...rcuit protection Power supply voltage must be sufficient to maintain a minimum of 9 VDC at the gauge terminals after dropping voltage across RL at full scale current 20 mA Example If RL 250 ohm then d...

Page 9: ...T PIN D MINUS 24 VDC POWER IN PIN E NO CONNECTION PIN F CASE GROUND 9 External Power Without Other Options The external power input is accessed through a miniature Phone Jack mate supplied on the back...

Page 10: ...DPG5000 M3366 0606 Notes...

Page 11: ...irectives OMEGA will add the CE mark to every appropriate device upon certification The information contained in this document is believed to be correct but OMEGA Engineering Inc accepts no liability...

Page 12: ...44 e mail nl omega com Czech Republic Rud arm dy 1868 733 01 Karvin TEL 420 0 69 6311899 FAX 420 0 69 6311114 Toll Free in Czech Rep 0800 1 66342 e mail czech omega com France 9 rue Denis Papin 78190...

Reviews: