Section 5 - Specifications
5.1 - General
Temperature Range
-18 to 538°C (0 to 1000°F)
Accuracy @ 22°C (72°F)
± 2% of Rdg. or 2.2°C (4°F) whichever is
ambient temperature &
greater
emissivity of 0.95 or greater
Optical Field of View
6:1 (Distance/Spot Size)
Repeatability
± 1% of Rdg.
Spectral Response
5 to 14 microns
Response Time
150 msec (0 to 63% of final value)
Emissivity Range
0.1 to 0.99, adjustable
Operating Ambient Temperature
Main Transmitter
0 to 50°C (32 to 122°F)
Sensor Head
0 to 70°C (32 to 158°F)
Sensor Head (-HT Model)
0 to 85°C (32 to 185°F)
Sensor Head with OS100-WC
(Water Cooling Jacket)
0 to 200°C (32 to 392°F)
Operating Relative Humidity
Less than 95% RH, non-condensing
Water Flow Rate for OS100-WC 0.25 GPM, room temperature
Thermal Shock
About 30 minutes for 25°C
abrupt ambient temperature change
Warm Up Period
3 minutes
Air Flow Rate for OS100-AP
1 CFM (0.5 Liters/sec.)
Power
12 to 24 VDC @ 100 mA
Analog Outputs
MV-F
1 mV/°F
MV-C
1 mV/°C
K
K Type TC - OS101 only
MA
4 to 20 mA
V1
0 to 5 VDC
Output Load requirements
Min. Load (0 to 5VDC)
1 K-Ohms
Max. Load (4 to 20 mA)
(Supply Power - 4 )/20 mA
Transmitter Housing
NEMA-4 & IP65, Die Cast Aluminum
Sensor Head Housing
NEMA-4 , Aluminum
Alarm Output
Open Drain, 100 mA
Alarm Set Point
0 to 100% , Adjustable via P4
Alarm Deadband
14°C (25°F)