Table
8
-1. Thermocouple Accuracy (ºC) for OM-DAQ-USB-2401
TC Type
Temp.
Typical
Fast
(°C)
Slow
-18
1.3
3.8
J
0
1.1
3
1200
0.9
2.5
-129
1.8
5
K
0
1.4
3.9
1372
1.3
3.7
-101
1.6
4.5
T
0
1.4
4
400
1
2.9
-184
1.2
3.4
E
0
0.9
2.6
1000
0.7
1.9
204
5.2
14.8
R
700
4.6
13.1
1768
3.9
11
204
5.6
16
S
700
5.2
14.7
1768
4.6
12.8
B
538
7.9
13.7
1820
4.8
13.7
-129
2.6
7.3
N
0
2
5.9
1300
1.4
4
Thermocouple accuracy excludes cold junction compensation
error.
Specifications
8
8
-3
NOTE: