SPECIFICATIONS

Display: 3½ digit liquid crystal display (LCD) with a maximum reading of 1999.
Polarity:  Automatic, positive implied, negative polarity indication.
Overrange: (OL) or (-OL) is displayed.
Zero: Automatic.
Low battery indication:  the "

" is displayed when the battery voltage drops

below the operating level.

Measurement rate: 2.5 times  per second, nominal.
Operating Environment: 0°C to 40°C at < 70% relative humidity.
Storage Temperature: -20°C to 60°C, 0 to 80% R.H. with battery removed

from meter.

Accuracy: Stated accuracy at 23°C ± 5°C, <75% relative humidity.
Safety: According to EN61010-1 protection class II overvoltage category

(CAT III 600V) pollution degree 2.

Clamp jaw: According to EN61010-2-032 CAT IV 600V.
Power: single standard 9-volt battery.
Battery life: 200 hours typical.
Dimensions: 250mm (H) x 100mm (W) x 46mm (D).
Weight: Approx. 380g including battery.
Accessories: One pair test leads, 9V battery.

Summary of Contents for HHM598F

Page 1: ...OMEGA HHM598F Digital Clamp On Meter...

Page 2: ...l Free in Germany 0800 82 66342 e mail nl omega com e mail germany omega com Czech Republic United Kingdom ISO 9002 Certified ul Rude armady 1868 733 01 Karvina One Omega Drive Hranice Czech Repubic R...

Page 3: ...working above 60V dc or 30V ac rms Such voltages pose a shock hazard 3 When Using the probes keep your fingers behind the finger guards on the probes 4 Measuring voltage which exceeds the limits of th...

Page 4: ...nominal Operating Environment 0 C to 40 C at 70 relative humidity Storage Temperature 20 C to 60 C 0 to 80 R H with battery removed from meter Accuracy Stated accuracy at 23 C 5 C 75 relative humidity...

Page 5: ...4dgts Input impedance 10MW Overload protection 600VDC or AC rms RESISTANCE CONTINUITY Ranges 2KW 200KW Accuracy 1 2 rdg 1dgt Open circuit volts 0 3Vdc Audible indication less than 30W on 2KW range Ove...

Page 6: ...ent 1 0mA 0 6mA Accuracy 6 0 rdg 3dgts Open circuit volts 3 0Vdc typical Audible indication 30mV Overload protection 600VDC or AC rms AC CURRENT Put conductor at the center of the jaws Ranges 20A 200A...

Page 7: ...out of DATA Hold mode In the DATA Hold mode the H annunciatorisdisplayed TheDATAHoldmodemaybeexited when changing function MAX HOLD Button Press MAX buttontotoggleinandoutofMAXHoldmode holdingthehighe...

Page 8: ...onto one conductor only Read the current directly on the display It is recommended that the conductor be placed at the center of the closed jaws for maximum accuracy 3 When the reading is lower than...

Page 9: ...Connect the test leads to the point of measurement and read the frequency from the display Temperature Measurements 1 Set the Function Range switch to the F position 2 Connect the temperature transit...

Page 10: ...6V typical for a silicon diode 5 Reverse probes If the diode is good OL is displayed If the diode is shorted 000 or another number is displayed 6 If the diode is open OL is displayed in both directio...

Page 11: ...transistor battery NEDA 1604 IEC 6F22 The appears on the LCD display when replacement is needed To replace the battery remove the two screws from the back of the meter and lift off the battery cover...

Page 12: ...ty or otherwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipmen...

Page 13: ...rs Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Conducti...

Reviews: