SETUP PARAMETERS 

TOTAL/BATCH TOTAL 
1.1 

SETUP - 1.1 determines the measurement unit for total and batch total The following units can 
be selected: 

L - m3 - UKGAL - USGAL - UKbbl - USbbl - OILbbl 

Please note that the K-Factor entered in 4.1 should be in pulses per litre and unit adjustment is 
done automatically. 

DECIMALS 
1.2 

The decimal point determines for total and batch total the number of digits following the decimal 
point. The following can be selected:  

 0000000  -  111111.1  -  22222.22  -  3333.333 

TOTAL ACCUMULATED 
2.1 

SETUP - 2.1 determines the measurement unit for accumulated total. The following units can be 
selected: 

L - m3 - UKGAL - USGAL - UKbbl - USbbl - OILbbl 

Please note that the K-Factor entered in 4.1 should be in pulses per litre and unit adjustment is 
done automatically. 

Decimals 
2.2 

The decimal point determines for accumulated total the number of digits following the decimal 
point. The following can be selected:  

 0000000  -  111111.1  -  22222.22  -  3333.333 

Flow Note 

The settings for total and flow rate are entirely separate. In this way, different units of measure-
ment can be used for each e.g. cubic meters for total and liters for flow rate. 
The display update time for flow rate is one second or more. 

RATE 
3.1 

SETUP - 21 determines the measurement unit for flow rate. 
The following units can be selected: 

 L - m3 - UKGAL - USGAL - UKbbl - USbbl - OILbbl 

Please note that the K-Factor entered in 4.1 should be in pulses per litre and unit adjustment is 
done automatically. 

TIME UNIT 
3.2 

The flow rate can be calculated per second (SEC), minute (MIN), hour (HR) or day (DAY). 

DECIMALS 
3.3 

This setting determines for flow rate the number of digits following the decimal point. The fol-
lowing can be selected:  

 00000   -  1111.1  -  2222.22   -  3333.333 

CALCULATION 
3.4 

The flow rate is calculated by measuring the time between a number of  pulses, for example 10 
pulses. The more pulses the more accurate the flow rate will be. The maximum value is 255 
pulses. 

Note:

 the lower the number of pulses, the higher the power consumption of the unit will 

be (important for battery powered applications). 

Note:

 for low frequency applications (below 10Hz): do not program more than 10 pulses 

else the update time will be very slow. 

Note:

 for high frequency application (above 1kHz) do program a value of 100 or more 

pulses. 

CUT-OFF TIME 
3.5 

With this setting, you determine a minimum flow requirement thresh-hold, if during this time 
less than XXX-pulses (SETUP 26) are generated, the flow rate will be displayed as zero. 
The cut-off time has to be entered in seconds - maximum time is 999 seconds (about 15 minutes). 

K-FACTOR
4.1

With the K-factor, the flow meter pulse signals are converted to a flow rate. 
The K-factor is based on the number of pulses generated by the flow meter per selected measure-
ment unit (SETUP 1.1), for example per liter. The more accurate the K-factor, the more accurate 
the functioning of the system will be.  
The setting allows a K Factor with up to 3 decimal places and 4 whole numbers eg; 

1234.123 

FPD3100-D DIGITAL REGISTER 

Set-up Functions 

Summary of Contents for FPD3100-D Series

Page 1: ...e mail info omega com For latest product manuals www omegamanual info FPD3100 D and FPD3100 D A Series 12mm LC Digital Display with Analog Output Options Shop online at omega com User sGuide...

Page 2: ...6 USA Canada only Tel 203 359 1660 Fax 203 359 7700 e mail info omega com For Other Locations Visit omega com worldwide omega com info omega com The information contained in this document is believed...

Page 3: ...ge 6 Wiring Diagram Page 7 Parts Listing Page 7 Technical Specifications Page 8 Type FPD3100 D A Digital Register Program Overview Page 10 Set up Functions Page 11 13 Wiring Diagram Page 13 Outputs Pa...

Page 4: ...polypropylene housing offers IP65 environmental protection Configuration of the unit The FPD3100 Series has been designed to be implemented in many types of applications For that reason a SETUP level...

Page 5: ...seconds Default 35 4 METER 4 METER SIGNAL Reed Switch Only 4 1 K FACTOR Pulses per Unit 0000 000 to 9999 999 5 OTHERS 6 1 TYPE DR 6 2 SOFTWARE VERSION 03 xx xx 6 3 SERIAL NO xxxxxxx GENERAL Configura...

Page 6: ...per litre and unit adjustment is done automatically TIME UNIT 3 2 The flow rate can be calculated per second SEC minute MIN hour HR or day DAY DECIMALS 3 3 This setting determines for flow rate the n...

Page 7: ...nput this sensor has been selected as the most common sensor and requires very little power with small effect on battery life The 2 position terminal block is not polarity conscious so the reed switch...

Page 8: ...A size Terminal connections Type Terminal strip Wire max 1mm Data protection Type Backup of all settings and running totals in flash memory Environment Electromagnetic compatibility Compliant ref EN...

Page 9: ...9 NOTES...

Page 10: ...m3 US GAL UK GAL US bbl UK bbl OIL bbl L 22 DECIMALS 0 1 2 3 0 FLOWRATE DEFAULT 31 UNIT L m3 US GAL UK GAL US bbl UK bbl OIL bbl L 32 TIME UNIT sec min hour min 33 DECIMALS 0 1 2 3 0 34 CALCULATION p...

Page 11: ...entered in 5 1 should be in pulses per litre and unit adjustment is done automatically TIME UNIT 3 2 The flow rate can be calculated per second SEC minute MIN hour HR or day DAY DECIMALS 3 3 This sett...

Page 12: ...endant upon SETUP 21 and 22 but are not displayed MAXIMUM FLOWRATE 6 2 Enter here the flow rate at which the output should generate the maximum signal 20mA or 10V in most applications at maximum flow...

Page 13: ...l be emptied It might be that pulses will be missed due to a buffer overflow so it is advised to program this setting within it s range PULSE PER 7 3 According to the measurement unit settings for tot...

Page 14: ...e output must be programmed as follows In menu 7 1 set 2222 22 In menu 7 2 set 10 or more 10mS or more if needed In menu 7 3 set 1 00 Example Pulse Alarm Output Setup Passive 4 20mA setup 9 10 Pulse O...

Page 15: ...15 FPD3100 D A DIGITAL REGISTER Parts Description...

Page 16: ...years 20 C ER17505 3 6V A size Terminal connections Type Terminal strip Wire max 1mm Data protection Type Backup of all settings and running totals in flash memory Environment Electromagnetic compati...

Page 17: ...the 4 00mA and 20 00mA levels precisely within set up Minimum operating voltage 8V DC Power supply 18 30V DC Switch outputs Function One scaled pulse output transmitting accumulated total One high fl...

Page 18: ...n 33 decimals 0000000 34 calculation pulses 010 35 cut off time 30 0 sec 4 Alarm 41 alarm disabled 42 flow zero on 43 alarm low 0 44 alarm high 9999 45 delay alarm 0 5 FLOWMETER 51 k factor 1 000 6 AN...

Page 19: ...PD3100 series is available with a Wall Mount Bracket option for applications requiring remote mount display Both the models D and D A are available with this option Following is a dimensional drawing...

Page 20: ...therwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Page 21: ...bine Paddlewheel Systems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U Industri...

Reviews: