IMPORTANT!
Use the type plate to check if the delivered unit corresponds to your order.
In particular, for devices with electrical components, check to see if the correct power
supply voltage is specified.
Accessories:
Connection cable with moulded M12x1 coupling
socket.
M12x1 coupling socket as component.
1.2 Intended use
The magnetic inductive flow sensor FMG70B must only be used for measuring and metering
liquids with a minimum conductivity of 50 μS/cm.
WARNING! No safety component!
The magnetic inductive flow sensor of the series FMG70B is not safety components in
accordance with Directive 2006-42-EC (Machine Directive).
Never use the FMG70B as a safety component.
The operational safety of the device supplied is only guaranteed by intended use. The
specified limits (
§ 9 "Technical data") may under no circumstances be exceeded.
Before installing the device, check that the wetted materials of the device are compatible with
the media being used (
Measuring tube empty (or partially filled). / Conductivity too low.
The green LED may blink irregularly if the measuring tube of the FMG70B is empty or
partially filled or if the conductivity of the fluid being used is too low. Random impulses
will be present at the output, but they do not represent an actual flow.
Ensure that the measuring tube of the FMG70B is always completely filled (
§ 4.1
Ensure that the conductivity of the fluid is at least 50 μS/cm.
1.3 Exclusion of liability
We accept no liability for any damage or malfunctions resulting from incorrect installation, in-
appropriate use of the device or failure to follow the instructions in this operating manual.
Summary of Contents for FMG70B Series
Page 1: ...FMG70B Series Magnetic Inductive Flow Sensor...
Page 2: ...2 M 5575 0416...
Page 21: ...For your notes Technical changes reserved 21 For your notes...
Page 22: ...For your notes 22 M 5575 0416 For your notes...
Page 23: ...Technical changes reserved 23...
Page 24: ...24 M 5575 0416 M 5575 0416...