background image

7

11. Fluid Filtering     

Heavily contaminated fluids or foreign matter in the fluid can block, da-
mage, and even destroy the flow meter. In these cases, always install a 
sufficiently large filter in front of the flow meter so that foreign particles and 

solids are prevented from entering the measuring elements, thus preventing 
damage to the flow meter. The required filtering depends on the size, bea-
ring, and design of the flow meter.

Flow meter size

Filter size for 

ball bearing

FHG 1xx2

250 µm

FHG 1xx4

250 µm

FHG 1xx5

500 µm

FHG 1xx7

500 µm

Table 2: Upstream filters

The filter size for flow meters with slide bearings, in special designs, or with  
specially adapted measuring element tolerances can be obtained from 
Omega upon request.

Important:

A blocked flow meter is capable of stopping the entire flow. 

An overpressure / bypass valve must be installed in the system side.

12. Sensor Electronics Function

The liquid to be measured flows through the rotor chambers in axial direc-
tion, resulting in an even rotation of the screw spindles.

This is done especially gentle and with very low resistance for the fluid to 
be measured as well as pulsation-free and almost free of leaks due to the 
specially designed fluidic profile geometry. 

A magnet wheel permanently affixed to the rotors is scanned without 
contact with a sensor module. The non-contact pick-up system consists of 
two GMR bridges (sin /cos), which are located in a sensor unit in cartridge 
design. It detects every movement of the sensing gear and routes the sin/
cos signals to the preamplifier electronics. The preamplifier electronics di-
gitise and amplify the sensor signals and multiply them by a high-resolution 
interpolator using adjustable settings. The square wave signals phase-shif-
ted by 90° are bidirectional and can be utilised by any evaluating device 
as well as computers and PLC controls. 

The flow is proportional to the edges/pulse count and the flow rate is pro-
portional to the frequency. The adjustable interpolator can be used to ad-
just the resolution explicitly to the downstream connected evaluating unit 
for obtaining highly precise measuring results of the entire system. This ap-
plies to the following application cases, for example:

• Measuring, controlling, and regulating high viscosity fluids
• Measuring, controlling, and regulating in lower flow ranges
• Measuring, controlling, and regulating when passing through zero
• Measuring, controlling, and regulating in both flow directions
• Measuring, controlling, metering, and filling of small volumes

The resolution is selectable in steps from factor 1 to 128. The frequency 
range is from 0 to 100 kHz.

In case of a 1-channel evaluation, a separate directional signal is availa-
ble.

The preamplifier is protected against reverse polarity and incorrect con-
nection. It is designed for fluid temperatures of -30°C up to +120°C and is 
mounted directly on the FHG flow meter.

The fluid volume passed through by one gear division of the sensing 
wheel within the measuring element is divided by the set interpolation fac-
tor.  
This forms the measurement volume per pulse (V

m

) with the defined unit 

[cm³/pulse].
The frequency of the output signals can be calculated as follows:

Formula 1: Calculation of the output frequency with Q in l/min

Table 2, Formula 2, and the subsequent diagrams can be used to determine 
the corresponding resolution or the corresponding IPF for the respective 
application.

Adjustable interpolation factors IPF: 1, 2, 5, 10, 25, 32, 50, 64, 100, 128 

f= Q x 1000

     V

m

     60

Summary of Contents for FHG 100

Page 1: ...omega com e mail info omega com For latest product manuals omegamanual info FHG FLOW METERS Shop online at User sGuide...

Page 2: ...ged by the United Kingdom Office Toll Free 0800 466 342 TEL 33 0 161 37 29 00 FAX 33 0 130 57 54 27 e mail sales omega fr Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany Toll Free 0 800...

Page 3: ...OPERATING INSTRUCTIONS For FHG Series Flow Meters...

Page 4: ...ics 11 15 Signaling LEDs 12 16 Operating Mode Messages 12 17 Alarm and Warning Messages 13 18 Preamplifier Technical Data 13 19 Preamplifier Pin Assignment 13 20 Maintenance 14 21 Returning for Repair...

Page 5: ...ply them by a high resolution interpolator using adjustable settings The square wave signals are bidirectional and can be utilised by any evaluating instrument as well as computers and PLC controls Th...

Page 6: ...tion check and verify the following properties aspects of the respective circumstances of your system to ensure operation is trouble free safe and reliable 1 The Fluid to be Processed Are the flow met...

Page 7: ...qualify as pressure equipment as defined by Section 1 Paragraph 2 1 4 of the directive listed above and as such are affected by the regulations of this directive Omega flow meters must therefore meet...

Page 8: ...Thread Please comply with the screw in depths and sealing systems PTFE tape or liquid sealants such as adhesives are not permitted Fastening The devices must be installed stress free into the pipeline...

Page 9: ...ply them by a high resolution interpolator using adjustable settings The square wave signals phase shif ted by 90 are bidirectional and can be utilised by any evaluating device as well as computers an...

Page 10: ...surement volume Vm cm3 Imp K Factor Imp l K Factor Imp gal 1 0 10 00000 100 379 2 1 5 00000 200 757 5 2 2 00000 500 1893 10 3 1 00000 1000 3785 25 4 0 40000 2500 9464 32 5 0 31200 3200 12113 50 6 0 20...

Page 11: ...9 FHG 1xx2 FLOW vs FREQUENCY FHG 1xx4 FLOW vs FREQUENCY Flow diagrams vs frequency...

Page 12: ...xample Flow meter FHG 1xx4 Max processable input frequency of the downstream evaluating unit 20 kHz Max operating flow 140 l min Path 1 The diagram yields an IPf of 25 Path 2 IPF fmax x VmIPF1 x 60 Qm...

Page 13: ...ser is able to set the degree of filtering in the form of partial volumes using rotary coding switches Adjustable pulse filtering 0Z 0 25Z 0 5Z 0 75Z 1 0Z 1 25Z 1 5Z 1 75Z 2 0Z 2 25Z 2 5Z 2 75 3 0Z 3...

Page 14: ...fly press the S2 key for the acknowledgment The new pulse rate is enabled a once The rotary coding switch for the pulse filtering has 16 switch settings The degree of filtering is determined with quar...

Page 15: ...tive flow 6 Temperature errors The temperature of the fluid is too high 120 C and may result in flawed or incorrect measurements 18 Preamplifier Technical Data Scanning sensor 2 x GMR sensors in a bri...

Page 16: ...ly well shielded connection cables with a wire cross section of 4 to 5 x 0 25 mm Please note that the housing of the round pin plug is metallic has a connection for the shield and that the potential o...

Page 17: ...flow meter has been completely cleaned and flushed This is to protect our employees and makes our work easier In case of noncompliance with this rule the devices are returned to the sender without att...

Page 18: ...sswiderstand p Durchflusswiderstand p Durchflussbereich 0 bis 50 l min Flow range 0 up to 120 l min Flow range Q Flow range Q Flow range Q Flow range Q Flow range Q Flow range pressure drop p Flow ran...

Page 19: ...sseinheit AR 2500 R Flanschanschlussma e DIN EN 1092 Gewicht 120 kg FHG1xx5 FHG1xx7 Preamplifier Preamplifier Earthing Earthing Weight 81 kg Weight 81 kg Weight 120 kg Sensor module Sensor module Conn...

Page 20: ...ier Preamplifier Earthing Earthing on both sides on both sides M8 15 deep M8 15 deep Weight 12 kg Weight 12 7 kg Weight 24 8 kg Weight 22 kg Sensor module Test Port Test Port Sensor module Connection...

Page 21: ...2 Evaluating unit e g display Flow sensor FHG Q 8l min V 11 53 l direction pos f 424 9 Hz control input 0 voltage blue channel 1 white channel 2 black Error Direction grey 25 Pin Assignment Channel 1...

Page 22: ...therwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Page 23: ...city Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industria...

Reviews: