![background image](http://html1.mh-extra.com/html/omega/dra-csm-1/dra-csm-1_quick-manual_3991253010.webp)
8
SUPPLY
Supply Voltage
Supply Current Consumption
FUSES
Main Fuse
Channel Fuse
TEMPERATURE
Operating
Storage
HUMIDITY
HOUSING
Box
Terminals
WEIGHT
DIMENSIONS
15 - 40 Vdc (regulated)
15mA in operation mode,
120mA in Test mode
(transmitters not included)
630mA, Fast Blow (20x5 mm)
63mA, Fast Blow (20x5 mm)
0 to 70
o
C (32 to 158
o
F)
-25 to +85
o
C (-13 to 185
o
F)
5 to 95% Relative humidity, non condensing
Plastic Polycarbonate
According to IP50 DIN 40050
According to IP20 DIN 40050
0.75 Kg. (1.5 lb.)
73Hx200Wx121mmD
(2.88"x7.88"x4.76")
INPUTS
ANALOG INPUTS
(Reverse Polarity Protected)
Max Input Current
CONTROL INPUTS
Logic Levels
Logic Polarity
OUTPUTS
CURRENT OUTPUT
Switching time
Accuracy (Refer to Current Input)
"V" TERMINALS
"V" Voltage
Limitation Current
Automatic Temperature Shut-down
INDICATORS
16, 0/4-20mA Current loop
30mA
4 Address inputs
E/T Enable/Test input
Low: "0" < 0.4V
High: 5 < "1" < 60V
Selectable - True High/True Low
4-20mA, Current Loop
10
m
Sec (into a resistive load)
0.01% maximum @ for the entire temperature range
V supply - 2.5V
40
±
1mA
Above 80
o
C
1 Yellow LED, Power-On indicator
16 Red LEDs, current activity indicators
9. SPECIFICATIONS