background image

DPG1000AD, DPG1001AD, DPG1000ADBL, DPG1100AD, DPG1100ADBL

2

Installation and Precautions

4

 Read these instructions before using the gauge. Contact 

the factory for assistance.

4

 These products do not contain user-serviceable parts. 

Contact us for repairs, service, or refurbishment.

4

 Gauges must be operated within specified ambient tem-

perature ranges.

4

 Outdoor or wash down applications require a NEMA 4X 

gauge or installation in a NEMA 4X housing.

4

 Use a pressure or vacuum range appropriate for the 

application.

4

 Use fittings appropriate for the pressure range of the 

gauge. 

4

 Due to the hardness of 316 stainless steel, it is recom-

mended that a thread sealant be used to ensure leak-free 
operation.

4

 For contaminated media use an appropriate screen or 

filter to keep debris out of gauge port. 

4

 Remove system pressures before removing or installing 

gauge.

4

 Install or remove gauge using a wrench on the hex fitting 

only. Do not attempt to turn gauge by forcing the housing. 

4

 Good design practice dictates that positive displacement 

liquid pumps include protection devices to prevent sensor 
damage from pressure spikes, acceleration head, and 
vacuum extremes.

Avoid permanent sensor damage! Do not apply vacuum to 
non-vacuum gauges or hydraulic vacuum to any gauges. 

Avoid permanent sensor damage! NEVER insert objects 
into gauge port or blow out with compressed air.

 Gauges are not for oxygen service. Accidental rupture of 

sensor diaphragm may cause silicone oil inside sensor to 
react with oxygen.

p

 NEVER connect the gauge wires directly to 115 VAC or 

permanent damage will result.

Power

The AD series is powered by 8-24 VAC 50/60 Hz or 9-32 
VDC.
The type and magnitude of the supply voltage have negli-
gible effects on the gauge calibration as long as it is within 
the voltage ranges stated above. No polarity needs to be 
observed when connecting a power supply. An inexpensive 
unregulated low voltage AC or DC power supply can be used.
After the gauge is installed, route the wires away from heat 
sources and moving equipment and connect the low-voltage 
power source to the gauge wires.
Ensure that the gauge supply voltage does not fall below 8 
VACRMS if AC power is used, or 9 VDC if DC power is used. 
Operation with less than these values may cause erratic or 
erroneous readings. 
When operating multiple gauges from the same power 
supply, refer to the mA rating in the specifications to ensure 
adequate power.
Note that standard 24 VAC transformers often operate at 
voltages well over the gauge’s 24 VAC limit. 

Absolute Reference Gauges

These models have engineering units shown on the rear 
label such as PSIA, INHGA, INH2OA, ZIN2A, BARA, MBARA, 
KPAA, TORR or TORRA, MMHGA, KGCMA, ATMA. The 
“A” indicates absolute reference. These ranges display 
atmospheric pressure when the gauge port is open to the 
atmosphere. They display zero at high vacuum. 
When the gauge port is open to the atmosphere, it is normal 
for the reading to constantly change in response to atmo-
spheric pressure changes. Do not adjust the calibration 
controls!

Display Backlighting (ADBL models)

Red LED display backlighting on ADBL models is on whenev-
er the gauge is on. Backlighting may not be apparent under 
bright lighting conditions.

Operation – Ranges up to 1999

Press the button on the front of the gauge to activate the 
display. 
The gauge can be shut off at any time by pressing the 
button again. 
If the gauge is in the power-on state and the power is dis-
connected, the gauge will turn on when power is reapplied.
The display indicates the pressure reading updated approxi-
mately 3 times per second. The gauge can be left on contin-
uously or turned off when not in use. 
Display backlighting for BL models is on whenever the gauge 
is on. The backlighting will not be apparent under bright 
lighting conditions.

Operation – Ranges 2000 and up including 
3000 psi, 5000 psi

Press and hold the front button for approximately 1 second if 
the gauge does not turn on when power is applied.
When the supply voltage is applied, the gauge will go 
through a power-up sequence. The full-scale range is indi-
cated, display segments are tested, and then the reading 
and units are displayed.
The gauge may be zeroed at power-up by following the 
procedure below. This feature corrects small deviations 
from zero due to temperature changes. Absolute reference 
gauges do not use the zero feature since they normally read 
atmospheric pressure.
The gauge port must be exposed to normal atmospheric 
pressure with no pressure applied. The zero function is only 
used at power-up and the stored zero correction is erased 
when the gauge is shut off. 
Press and hold the front button.
The full-scale range is indicated and the display is tested. 
Continue to press the button until 

oooo

 is displayed and 

then release the button.
The gauge is now zeroed and ready for use with the actual 
pressure is displayed.
Attempting to zero the gauge with pressure greater than 
approximately 3% of full-scale applied will result in an error 
condition, and the display will alternately indicate 

Err0

 and 

the actual measured pressure. The gauge must be powered 
down to reset the error condition.
Following the start-up initialization, the display indicates the 
pressure reading updated approximately 3 times per second. 
If excessive vacuum is applied to a pressure-only gauge, 
the display will indicate 

–Err

 until the vacuum is released. 

Applying vacuum to a gauge designed for pressure may 
damage the pressure sensor. If excessive pressure is 
applied (112.5% over range), an out-of-range indication of 

I – – –

 or 

I –.–.–.–

 will be displayed depending on model.

Display backlighting for BL models is on whenever the gauge 
is on. The backlighting will not be apparent under bright 
lighting conditions.
To shut off the gauge at any time, press and hold the button 
until the display indicates 

OFF

 (about 5 seconds) and then 

release.

CAUTION:

1

/

4

”NPT

3.38”

0.75”

PSIG

Span

Zero

2.88”

1.65”

Turn at hex 

fitting only!

1000AD, 1001AD, 1000ADBL

OMEGA Engineering Inc./Made in USA

3

1

/

2

 digit or  

4

1

/

2

 digit display

1

/

4

”NPT

3.0”

3.5”

2.0”

Turn at hex 

fitting only!

PSIG

Span

Zero

OMEGA Engineering Inc. / Made in USA

0.75”

1100AD and 1100ADBL (NEMA 4X)

3

1

/

2

 digit or  

4

1

/

2

 digit display

Allow 1” 

clearance

3 ft long, 
2-conductor, 
22 AWG cable

Allow 1” 

clearance

3 ft long, 
2-conductor, 
22 AWG cable

WARNING:

 This product can expose you to chemicals including 

lead, nickel and chromium, which are known to the State of 

California to cause cancer or birth defects or other reproductive 

harm. For more information go to www.P65Warnings.ca.gov

Reviews: