![Omega DP6060 User Manual Download Page 70](http://html1.mh-extra.com/html/omega/dp6060/dp6060_user-manual_4062237070.webp)
DP6060 Dual-Input Process Meter
Instruction Manual
70
Advanced Features Menu & Display Messages
The following table shows the functions and messages of the
Advanced
Features
menu in the order they appear in the menu.
Display Parameter Action/Setting
filtEr
Filter
Set noise filter value
Ch-A
Channel A
Set filter value for channel A
Ch-b
Channel B
Set filter value for channel B
bypASs
Bypass
Set filter bypass value
Ch-A
Channel A
Set filter bypass value for channel A
Ch-b
Channel B
Set filter bypass value for channel B
Round
Round
Set the rounding value for display
variables
seriAl
Serial
Set serial communication parameters
SlaveId
Slave ID
Set slave ID or meter address
Baud
Baud rate
Select baud rate
Tr dLY
Transmit delay
Set transmit delay for serial
communication
Parity
Parity
Select parity
Even, Odd, or None with 1 or 2 stop bits
t-byt
Time byte
Set byte-to-byte timeout
SelEct
Select
Enter the Select menu (function, cutoff,
out)
Functn
Signal input
conditioning
Select linear, square root, programmable
exponent, or round horizontal tank function
Ch-A
Channel A
Select menu for channel A
Ch-b
Channel B
Select menu for channel B
Linear
Linear
Set meter for linear function and select
number of linearization points
No pts
Number of
points
Set the number of linearization points
(default: 2)
Square
Square root
Set meter for square root extraction
Prog E
Programmable
exponent
Set meter for programmable exponent and
enter exponent value
Rht
Round
horizontal tank
Set meter for round horizontal tank volume
calculation
Length
Length
Enter the tank’s length in inches
dianmr
Diameter
Enter the tank’s diameter in inches
nmath
Math
Enter the setup menu for channel C math
functions
Sunm
Sum
Channel C = (A+B+P)*F
Dif
Difference
Channel C = (A-B+P)*F
difabs
Absolute
difference
Channel C = ((Absolute value of (A-
B))+P)*F
Avg
Average
Channel C = (((A+B)/2)+P)*F
nmulti
Multiplication
Channel C = ((A*B)+P)*F