![background image](http://html1.mh-extra.com/html/omega/dp3001/dp3001_user-manual_4045886016.webp)
10
5.2
VOLTAGE RANGE SELECTION
Figure 5.2 Main Board Jumper Locations
*Refer to
Figure 5.3.
S3-A is used for signal connections. Refer to
Section 4 for more information.
Input
Configuration
Solder Switches*
Push-on Jumpers Wire Jumper
Open
Close
S2
S4
W10
2 Vdc
J, L
H, I, K
-
-
Install
20 Vdc
J, L
H, I, K
A
B
Remove
200 Vdc
J, L
H, I, K
A
C
Remove
Summary of Contents for DP3001
Page 2: ......
Page 6: ......
Page 18: ...12 7 0 DRAWINGS Figure 7 1 Assembly Diagram Main and Display Board...