background image

8

4.4 

SIGNAL INPUTS

RATIO:  The reference input allows an external voltage to be used as the 

reference source for conversion. In this mode, the meter reads the ratio of the 

signal voltage to the reference voltage rather than the true value of the input.

Signal Voltage 

Reading in Counts =

     Reference Voltage  

x 10000

On the 20 V and 200 V ranges, the signal voltage must be scaled by 1/10 and 

1/100, respectively. For all ranges, the standard reference input impedance ratio 

is 80 ohm. For 100 Mohm, open solder switch F on the main board (

Figure 5.3). 

The reference voltage must be between the limits specified, +0.5 to +2.0 V, and 

must be positive with respect to analog ground.

HOLD:  When high (or open), the A/D will free-run with equally spaced 

measurement cycles every 40,002 clock pulses. If taken low, the converter will 

complete the full measurement cycle and then hold this reading as long as HOLD 

is low. A positive pulse (greater than 300 ns) will now initiate a new measurement 

cycle, beginning with between 9,001 and 10,001 counts of auto-zero time. If the 

pulse terminates before the full measurement cycle (40,002 counts) is complete, 

it will not be recognized and the converter will simply complete the present 

measurement. An external indication that a full measurement cycle has been 

completed is that the first strobe pulse will occur 101 counts after the end of 

this cycle. Thus, if HOLD has been low for at least 101 counts, the converter is 

holding and ready to start a new measurement when pulsed high.

BLANKING: The display may be blanked by grounding the BLANKING input. 

The blanking input must be open for normal display operation. The polarity sign 

and decimal points are not blanked, but they will flash if the displayed reading 

exceeds ±19999 counts.

Summary of Contents for DP3001

Page 1: ...http www omega com e mail info omega com User sGuide DP3001 DIGITAL PANEL VOLTMETER ROHS 2 COMPLIANT...

Page 2: ......

Page 3: ...nversion 2 2 7 Display 2 2 8 Power 2 2 9 Environmental 2 2 10 Mechanical 3 3 0 MECHANICAL ASSEMBLY INSTALLATION 3 3 1 Case Dimensions 3 3 2 Panel Mounting 4 4 0 POWER SIGNAL 5 4 1 Changing Operating V...

Page 4: ...1 Side View of Transformer 6 Figure 4 2 Rear Terminal Hookups 7 Figure 5 1 Display Board Jumper Locations 10 Figure 5 2 Main Board Jumper Locations 11 Figure 5 3 Solder Switch Locations 12 Figure 7 1...

Page 5: ...incorporated in the mains supply cord Furthermore to provide protection against excessive energy being drawn from the mains supply in case of a fault in the equipment an over current protection device...

Page 6: ......

Page 7: ...del contains the required circuitry for full range selection and easy configuration 2 0 SPECIFICATIONS 2 1 ANALOG INPUT Model 3 4 5 Range 1 9999 Vdc 19 999 Vdc 199 99 Vdc Resolution 0 1 mV 1 mV 10 mV...

Page 8: ...ogrammable internally or at the rear connector Overrange indication 4 least significant digits flash 2 8 POWER Standard AC power voltage 115 or 230 Vac 15 AC frequency range 49 to 440 Hz Optional DC p...

Page 9: ...ouble row of 18 pins 3 96 mm 0 156 in spacing between pins Barrier terminal strip with six 6 screw Bezel Depth behind bezel with connector Panel cutout Weight Case material D1 connector option D4 conn...

Page 10: ...the panel cutout refer to case dimensions shown below 3 Push the slide retainers onto the case to fit tightly against the panel rear Rotate slide clamps back into original position and tighten clamp s...

Page 11: ...with the shield terminated to analog ground at the connector Both analog and digital ground are internally connected and should not be connected externally For proper grounding connect the low signal...

Page 12: ...d Green Green Model Power Signal AC D1 Connector A High 17 High 2 Low 16 Low C Earth Ground T Analog Ground D4 Connector 1 High 6 High 2 Low 5 Low 3 Earth Ground 4 Analog Ground DC D1 Connector 2 DC 1...

Page 13: ...E24 H 199 99 DP3 7 Spare E23 J 19 999 DP2 8 Spare E22 K 1 9999 DP1 9 Spare E25 L Decimal point select 10 Spare E26 M Spare E27 11 4 6 Vdc power output N Spare E29 12 4 7 Vdc power output P Excitation...

Page 14: ...s If taken low the converter will complete the full measurement cycle and then hold this reading as long as HOLD is low A positive pulse greater than 300 ns will now initiate a new measurement cycle b...

Page 15: ...lder switches as indicated Remove all push on jumpers not used 5 1 DECIMAL POINT SELECTION Figure 5 1 Display Board Jumper Locations Decimal Point S1 Alternate decimal point configuration using main b...

Page 16: ...ons Refer to Figure 5 3 S3 A is used for signal connections Refer to Section 4 for more information Input Configuration Solder Switches Push on Jumpers Wire Jumper Open Close S2 S4 W10 2 Vdc J L H I K...

Page 17: ...the stability and internal accuracy If calibration is needed use the following procedure 1 Remove the front lens Insert a blade screwdriver under the notch at the bottom of the lens and gently pry it...

Page 18: ...12 7 0 DRAWINGS Figure 7 1 Assembly Diagram Main and Display Board...

Page 19: ...t the slope in volts sec during this second period T2 is constant and independent of input signal levels The time required to discharge the capacitor back to zero volts is then proportional to the sig...

Page 20: ..._______________________ _____________________________________________________________________________ ______________________________________________________________________________ ___________________...

Page 21: ..._________________________ _____________________________________________________________________________ ______________________________________________________________________________ _________________...

Page 22: ..._________________________ _____________________________________________________________________________ ______________________________________________________________________________ _________________...

Page 23: ...NS Equipment sold by OMEGA is not intended to be used nor shall it be used 1 as a Basic Component under 10 CFR 21 NRC used in or with any nuclear installation or activity or 2 in medical applications...

Page 24: ...Indicators R Turbine Paddlewheel Systems R Totalizers Batch Controllers pH CONDUCTIVITY R pH Electrodes Testers Accessories R Benchtop Laboratory Meters R Controllers Calibrators Simulators Pumps R In...

Reviews: