This product conforms to the EMC directive 89/336/EEC amended by
93/68/EEC, and with the European Low Voltage Directive 72/23/EEC.
Electrical Safety EN61010-1:2001
Safety requirements for electrical equipment for measurement, control
and laboratory.
•
Double Insulation: Primary to Secondaries
•
Pollution Degree 2
•
Measurement *Category I
Note 1:
*Measurements performed on circuits not directly connected
to the Mains Supply (power). Maximum Line-to-Neutral working
voltage 50Vac/dc.
Note 2:
This unit should not be used in Measurement Categories II, III, IV.
•
Dielectric withstand Test per 1 min (50/60Hz)
Power to Input/Output 2500Vac
•
Transients Overvoltage
Input Power Transients Overvoltage 2500V
Input/Output Signals Transients Overvoltage 500V
EMC EN61326:1997 + and A1: 1998 + A2: 2001
Immunity and Emissions requirements for electrical equipment for
measurement, control and laboratory.
•
EMC Emissions Class B, EN61326
•
EMC Immunity** Table 1, EN61326
** I/O Signal and Control Lines require shielded cables and these cables
must be located on conductive cable trays or in conduits. Furthermore,
the length of these cables should not exceed 30 meters.
Refer to the EMC and Safety installation considerations (Guidelines)