2.4.5 Calibration Menu
Caution: It is not necessary to calibrate a
brand new meter, it arrives completely
calibrated. Accessing the calibration
menu will affect the meter’s calibration
and should only be performed by
qualified personnel with accurate test
equipment.
2.4.5.1 Meter SPAN calibration
procedure
This item will only be available if the S4
jumper, which enables the calibration
configuration, is installed!
The Span calibration does not require any
special compensated thermocouple wires at
the input terminals. Standard copper wire
hooked up to a mV calibration source is
required.