SENSITIVITY PROGRAMMING:
Pick a sensitivity setting that is as close as possible, without going under, the expected
full-scale output from the transducer being used.
Example: Assume that a 10,000 pound capacity load cell with an output of .5mV/V
full-scale is going to be used with the DP-7600. Also assume that the excitation
required 10V; therefore, the full-scale input to the unit will be 5mV at 10,000 pound,
and the sensitivity switches located on the rear of the DP-7600 should be set for 5.1mV
full-scale (S1 - S3 open).
DP-7600 BACK VIEW
EXCITATION
SETTINGS
DP-7600 SENSITIVITY SETTINGS
14