![Omega DP-7600 User Manual Download Page 11](http://html2.mh-extra.com/html/omega/dp-7600/dp-7600_user-manual_5154771011.webp)
LEGEND SETUP
The DP-7600 can append a user selectable engineering units string to read-
ings transmitted from the unit. This feature allows direct terminal readout in engineer-
ing units, or easier unit identification in multi-unit applications. Program the legend
number to the number corresponding with desired units as shown in the table below.
Legend Table
0 - No Legend 4 - kpa.
1 - lbs. 5 - mV.
2 - kgs. 6 - mA.
3 - psi. 7 - V.
CONTINUOUS UPDATE
The model DP-7600 can be programmed to automatically transmit the dis-
play reading at selectable time intervals. The update interval can be programmed from
20 times per second (-1) up to a maximum of 3600 seconds (3600). A value of zero
disables the automatic update function.
DEVICE CODE
A non-printable device code allows units to take turns on a serial bus when
daisy chained to a printer or some other data recording device. A value of one enables
transmission of the device code following the reading string; conversely, a value of
zero disables transmission of the device code.
SERIAL COMMAND
A serial command can automatically be executed upon receipt of a non-
printable code (see device code above). A table of numbers associated with each serial
command can be found in the serial communications section of the manual (on page
8). Program to the number associated with the command to be executed when the unit
receives a device code character on its serial communications input.
6
Summary of Contents for DP-7600
Page 21: ...SCHEMATIC DP 7600 Top Board 16...
Page 22: ...SCHEMATIC DP 7600 Top Board 17...
Page 23: ...SCHEMATIC DP 7600 Top Board 18...
Page 24: ...SCHEMATIC DP 7600 Bottom Board 19...
Page 25: ...SCHEMATIC DP 7600 Bottom Board 20...
Page 26: ...SCHEMATIC DP 7600 Analog Input Option Board 21...
Page 27: ...SCHEMATIC DP 7600 Output Option Board 22...
Page 28: ...SCHEMATIC DP 7600 RS232 RS485 Option Board 23...