Page 4

SOURCE

Choose  this  function  to  provide  a  simulated  thermocouple  signal  into  controllers,  temperature 
transmitters, indicators or any input devices that measure thermocouple sensors..

1)  Disconnect  the  thermocouple  sensor  from  the 

device to be calibrated.

2)  Select “

SOURCE

” with slide switch 

.

3)  Connect a thermocouple wire (matching the type 

of wire to sensor being simulated) with miniature 
male T/C  connector  to  the  inputs  of  the  device 
being  calibrated,  making  sure  to  check  polarity. 
Millivolt  outputs  (without  cold  junction)  may  be 
connected  with  a  copper  (white)  miniature  T/C 
connector or the banana jacks with copper wire.

The  output  is  adjusted  in  0.1°  (or  0.001  mV) 
increments  by  turning  the  knob 

while  the 

DIAL  switch 

 is  in  the  “HI”,  “LO”  or  “SET”  

position.  Press  and  turn  the  knob  for  faster  dialing 
with 10° (or 0.100) increments.

Sourcing Thermocouple

Storing DIAL Outputs

STORING HI and LO DIAL Outputs

Choose  this  function  to  provide  a  simulated  thermocouple  signal  into  controllers,  temperature 
transmitters, indicators or any other input device that measure thermocouple sensors.

1)  Store your high (SPAN) output temperature by moving the DIAL switch to the 

HI

 position  

and turn the DIAL knob until the desired temperature is on the display. Press and hold the  
DIAL knob until 

STORED

 appears to store the value. Release the DIAL knob.

2)  Store your low (ZERO) output temperature by moving the DIAL switch to the 

LO 

position

and turn the 

DIAL knob until the desired temperature is on the display. Press and hold the 

DIAL knob until 

STORED

 appears to store the value. Release the DIAL knob.

3)  Instantly output your SPAN and ZERO temperature outputs by moving the DIAL switch 

between HI and LO. You may also select any third temperature output (such as mid-range) using 
the SET position on the DIAL switch.

HI

TYPE K °C

HI

TYPE K °C

OFF

Push & Hold

to Store

Push & Turn

for

Fast Dialing

Double Click

for Options

THERMOCOUPLE SOURCE

SOURCE

HI

SET

LO

TYPES

322.1

(Optional Boot Shown)

Page 5

General

)

V

m

 

9

0

0

.

0

 

+

 

g

n

i

d

a

e

R

 f

o

 

%

5

1

0

.

0

(

±

y

c

a

r

u

c

c

A
Cold Junction Compensation

± 0.45°F (±0.25°C)

V

m

 

0

0

0

.

0

8

 

o

t

 

0

0

0

.

3

1

-

e

g

n

a

R

 

tl

o

v

il

li

M
Operating Temperature Range

-25 to 60°C (-10 to 140°F)

Temperature Drift

 50 ppm of range (includes mV and Cold Junction)

Relative Humidity Range

10%  RH  90% (0 to 35°C), Non-condensing
10%  RH  70% (35 to 60°C), Non-condensing

Size
       With Boot

4.96 x 2.73 x 1.79 inches, 126 x 69 x 45 mm (L x W x H)
5.67 x 3.06 x 2.05 inches, 144 x 78 x 52 mm (L x W x H)

Weight
      With Boot

8.4 ounces, 0.24 kg (including batteries)
11 ounces, 0.32 kg (including batteries)

)

6

R

L

(

 

V

5

.

1

 

e

n

il

a

k

l

A

 

A

A

 

r

u

o

F

s

e

i

r

e

t

t

a

B

s

r

u

o

H

 

0

5

e

fi

L

 

y

r

e

t

t

a

B
Optional NiMh Rechargeable battery kit

120 VAC for North America Only; charger,  four NiMh 

batteries, AC & DC cords [Part # 020-0103]

t

f

e

e

fi

f

o

 

r

u

o

h

 

1

 l

a

n

i

m

o

n

 

h

ti

w

 

n

o

it

a

c

i

d

n

y

r

e

t

t

a

b

 

w

o

L

y

r

e

t

t

a

B

 

w

o

L
Protection against misconnection

Over-voltage protection to 60 V dc (rated for 30 seconds)

 

g

n

it

h

g

il

k

c

a

b

 

D

E

L

 .

y

a

l

p

s

i

d

 l

a

t

s

y

r

c

 

d

i

u

q

il

 

c

i

h

p

a

r

g

 

t

s

a

r

t

n

o

c

 

h

g

i

H

y

a

l

p

s

i

D

for use in low lit areas.

Source

s

m

h

O

 

3

.

0

 

e

c

n

a

d

e

p

m

t

u

p

t

u

O

)

s

m

h

O

 

0

1

 

o

t

n

V

m

 

0

8

 

s

e

v

i

r

d

(

 

A

m

 

0

2

 

t

n

e

r

r

u

C

 

e

c

r

u

o

S
Noise

 4 microvolts p-p for frequencies of 10 Hz or below

Specifications

(Unless otherwise indicated all specifications are rated from a nominal 23 °C,

70 % RH for 1 year from calibration)

Additional Information 

This product is calibrated on equipment traceable to NIST and includes a Certificate of Calibration. 
OMEGA Engineering recommends a calibration interval of one year.

J

T

E

K

R

S

B

N

G

C

D

L

U

P

mV

1

CL510AZ

Reviews: