Page 6
T/C
Type
Degrees C
Range
Accuracy
Degrees F
Range
Accuracy
T/C
Material
ISA/ANSI
Color
J
-200.0 to -180.0
±0.5°
-346.0 to -292.0
±0.9°
+Iron
-Connstantan
Jacket
White
Red
Black
-180.0 to -50.0
±0.4°
-292.0 to -58.0
±0.7°
-50.0 to 500.0
±0.3°
-58.0 to 932.0
±0.5°
500.0 to 1200,0
±0.4°
932.0 to 2192.0
±0..7°
K
-230.0 to -100.0
±0.8°
-382.0 to -148.0
±1.4°
+ Chromel®
-Alumel®
Jacket
Yellow
Red
Yellow
-100.0 to 1050.0
±0.4°
-148.0 to 1922.0
±0.7°
1050.0 to 1371.1
±0.5°
1922.0 to 2500.0
±0.9°
T
-260.0 to -200.0
±1.2°
-436.0 to -328.0
±2.2°
+Copper
-Constantan
Jacket
Blue
Red
Blue
-200.0 to -50.0
±0.7°
-328.0 to -58.0
±1.3°
-50.0 to 0.0
±0.4°
-58.0 to 32.0
±0.7°
0.0 to 400.0
±0.3°
32.0 to 752.0
±0.5°
E
-240.0 to -200.0
±0.6°
-400.0 to -328.0
±1.1°
+Chromel
-Constantan
Jacket
Purple
Red
Purple
-200.0 to -100.0
±0.4°
-328.0 to -148.0
±0.7°
-100.0 to 850.0
±0.3°
-148.0 to 1562.0
±0.5°
850.0 to 1000.0
±0.4°
1562.0 to 1832.0
±0.7°
R
-13.3 to 250.0
±1.4°
-1.0 to 482.0
±2.5°
+Pt/13Rh
-Platinum
Jacket
Black
Red
Green
250.0 to 750.0
±0.8°
482.0 to 1382.0
±1.4°
750.0 to 1600.0
±0.7°
1382.0 to 2192.0
±1.3°
1600.0 to 1767.8
±0.8°
2192.0 to 3214.0
±1.4°
S
-18.3 to 100.0
±1.4°
-1.0 to 212.0
±2.5°
+Pt/10Rh
-Platinum
Jackrt
Black
Red
Green
100.0 to 400.0
±1.0°
212.0 to 752.0
±1.8°
400.0 to 1700.0
±0.8°
752.0 to 3092.0
±1.4°
1700.0 to 1767.8
±0.9°
3092.0 to 3214.0
±1.6°
B
315.6 to 550.0
±2.0°
600 to 1022.0
±3.6°
+Pt/30Rh
-Pt/6Rh
Jacket
Grey
Red
Grey
550.0 to 900.0
±1.3°
1022.0 to 1652.0
±2.3°
900.0 to 1150.0
±0.9°
1652.0 to 2102.0
±1.6°
1150.0 to 1820.0
±0.8°
2102.0 to 3308.0
±1.5°
Ranges & Accuracies
(Cold Junction Accuracy not included)
Page 3
Configuration
Configure the Calibrator
Move
POWER SWITCH to
“SOURCE”
or
“READ”
.
Setup
Double click the
DIAL KNOB at any time the
unit is on and the following displays will appear
for 15 seconds:
Turn the
DIAL KNOB to move through the
menu. Press the
DIAL KNOB to toggle
between OFF and ON or to scroll through the
settings.
MODEL 51#B V#.##
DOUBLE CLICK
DIAL KNOB
FOR CONFIGURATION
> EXIT
15
TEMP UNITS °C
T/C
K
AUTO OFF
ON
EXIT MENU
- exits this menu immediately and
saves any changes. Menu will automatically exit
after 15 seconds of inactivity (countdown timer is
displayed).
TEMP UNITS
- pressing the knob will toggle
between °C and °F.
T/C
-
520: pressing the knob will toggle between the
factory configured T/C Type and mV.
521: pressing the knob will cycle through T/C
types J, T, E, K, R, S, B, N, G, C, D, L (J-DIN), U
(T-DIN), Platinel II and mV.
AUTO OFF
- If AUTO OFF is ON, the unit will
turn off after 30 minutes of inactivity to save
battery life. If AUTO OFF is OFF the unit will stay
on until the POWER SWITCH is moved to the off
position.
Note:
All settings are remembered even with the
power off. Removing the batteries resets the
values to factory defaults.
Simulating thermocouples requires the use of
thermocouple or extension grade thermocouple
wire.
Plug thermocouple wires into the female
miniature thermocouple connector mounted in
the top end of the housing.
The OMEGA CL540A has two banana jacks
mounted in the top end of the housing. These
are not temperature compensated and are to be
used only for millivolt signals.
Connections
Push & Hold
to
Store/Ramp
Push & Turn
for
Fast Dialing
Double Click
for Options
HI
TYPE K °C
Thermocouple Wire
Device
Requiring
Calibration
HI
TYPE K °
C
OFF
THERMOCOUPLE SOURCE
SOURCE
HI
SET
LO
J
•
T
•
E
•
K
•
R
•
S
•
B
•
N
•
G
•
C
•
D
•
L
•
U
•
P
•
mV
TYPES
322.1
(Optional Boot Shown)
CL510AZ