Storing HI/SET/LO Outputs
STORING HI and LO Outputs
Choose this function to select commonly used temperature output values, and set high and low
values for stepping.
1) Store your high (SPAN) output temperature by moving the HI/SET/LO switch to the
HI position
and turn the
e
knob until the desired temperature is on the display. Press and hold the knob
until
STORED appears to store the value. Release the knob.
2) Store your low (ZERO) output temperature by moving the HI/SET/LO switch to the
LO position
and turn the
e
knob until the desired temperature is on the display. Press and hold the knob
until
STORED appears to store the value. Release the knob.
3) Instantly output your SPAN and ZERO temperature outputs by moving the HI/SET/LO switch
between HI and LO. You may also select any third temperature output (such as mid-range) using
the SET position on the HI/SET/LO switch.