17
7. Specifications
Functions:
mA source, mA simulate, mA read, mA read/
loop power, and volts read.
Ranges:
mA (0 to 24mA) and Volts (0 to 30VDC)
Resolution:
1uA on mA ranges and 1mV on voltage range
Accuracy:
0.01% +/- 2LSD all ranges (@23° +/- 5°C)
Operating Temp Range: -10°C to 55°C
Humidity Range:
10 to 95% non-condensing
Stability:
20ppm of F.S. /°C from -10°C to 18°C and 28°C
to 55°C
Display:
128 x 64 pixels, LCD Graphic w/backlight, .34”
high digits
Power:
6AAA alkaline, lithium, or NiMH batteries
Battery Life:
≥
40 hours continuous use (measure mode)
Loop Compliance
Voltage:
24VDC @ 20mA
Loop Drive Capability:
1200
Ω
without HART resistor, 950
Ω
with HART
resistor
Over-Voltage Protection: 240VAC
Overload Current
Protection:
28mA DC
EMC:
EN61326 Annex A (Portable Instruments)
Dimensions (LxWxD):
6” x 3.6” x 1.3” (15 cm x 9 cm x 3 cm)
Weight:
9.5 ounces (0.3 kg)
Included Accessories:
Traceable calibration certificate with data,
batteries, test leads, and manual